ESPECTROFOTÓMETRO UV/VIS 4320/1 - 4320/2 4320/1 - 4320/2 UV/VIS SPECTROPHOTOMETER SPECTROPHOTOMÈTRE UV/VIS 4320/1 - 4320/2 Este manual es parte inseparable del aparato por lo que debe estar disponible a todos los usuarios del equipo. Le recomendamos leer atentamente el presente manual y seguir rigurosamente los procedimientos de uso para obtener las máximas prestaciones y una...
ENGLISH LANGUAGE INDEX 1-30 Spanish 32-60 English 62-90 French TABLE OF CONTENTS Safety Information ........................34 Package Contents ........................34 Unpacking..........................34 Installation ..........................35 Symbols and Conventions ....................... 35 Overview ..........................35 Specifications ........................36 Description ..........................37 Getting Started ........................38 General Operating Instructions ....................
ENGLISH SAFETY INFORMATION Please follow the guidelines below and read this manual in its entirety to ensure safe operation of the unit. Do not open the device. Disconnect the device from the mains supply before carrying out maintenance work or changing the fuses.
ENGLISH INSTALLATION Placement Place the instrument on the stable table carefully. Install printer (Optional) Check to confirm instrument power switch is turned off; connect the printer’s data cable to the Instru- ment’s serial/USB port. Information: The spectrophotometer supports USB printers using the HP PCL3 GUI print description language.
ENGLISH GETTING STARTED The following chart describes the basic operation of the instrument. Turn On and Self-check Switch on the power. Self-check includes the following steps: Turn On Lamp - Locating Filter Disc - Lo- cating Automatic Sample Holder (If Installed) - Get Dark Current - Locating Wavelength - Check Energy. Page 38 Version 1, April 2024 User’s manual for UV/Vis Spectrophotometer...
ENGLISH GENERAL OPERATING INSTRUCTIONS Touch Screen Using Tips The entire screen can be started with a touch. To make a choice, use your nails, fingertips, pencil, or stylus to press the screen. Don’t press the screen with sharp objects (such as ball point). Select Application Main Interface, press the icon to select application.
ENGLISH Basic Operation Home Back to main interface. Return Back to the previous interface. Page Up/Down Go to previous/next page. Measurement Results Operation Open Open result(s) from internal/USB memory. Save Save result(s) to internal/USB memory. Print Print result(s). Delete Delete selected result(s). Rename, Print and Delete Results Rename a Sample: List interface, press the area Name, key in the sample name (up to 8 characters).
Page 11
ENGLISH Open Results Open: 1. List interface, press the icon 2. Press the icon internal memory/USB memory to select the memory which the file saved. 3. Press file lists to select, press the button Open. Save Results Save: 1. List interface, press the icon Save. 2.
ENGLISH Files Operation Internal Memory Internal memory of the spectrophotometer. USB Memory USB extended mass memory. Copy Copy the selected file(s) from internal /USB memory to USB/internal memory. Export csv Export file(s) to *.csv format Export txt Export file(s) to *.txt format Delete Delete the selected file(s).
ENGLISH CALIBRATION AND SYSTEM SETTINGS Select the icon in the main interface. Display options to calibrate the system and configure the basic instrument settings. Calibrate Start to do calibration. Calibration Select Tab Calibration in the System interface. Remove something in the measurement channel, close the sample chamber cover, select the item Dark current or Wavelength, press the icon to do calibration.
ENGLISH Change the lamp switching point: Press the value of lamp switching point. Input the new value. Reset the lamp usage: Press the icon to reset the Tungsten lamp/Deuterium lamp usage time. Important information: If only one of the light sources is used in for a long period of time, please turn off another light source to save energy.
ENGLISH Accept Accept the new language. Language Selection Select Tab Language in the System interface. Select a language, press the icon to change. General Options Select Tab General Options in the System interface. Beep: Press the icon to turn on/off the beep. Brightness: Press the icon to decrease/increase the brightness of the LCD display.
ENGLISH Restore Defaults Restore Restore the parameters to factory settings. Select Tab Restore defaults in the System interface. Select an item, press the icon to restore. PERFORMANCE VERIFICATION Select the icon in the main interface. Display options to verify the performance of the instrument. Important information Before verifying the performance, the instrument needs to be preheated for 30 minutes, and then re-measure dark current.
ENGLISH Verifying Wavelength Accuracy and Wavelength Repeatability Select Tab Wavelength accuracy in the Performance verification interface. Standard Sample: Holmium oxide solution or equivalent filter Measurement: 1. Remove something in the measurement channel, close the sample chamber cover, press the wavelength value, type in the wavelength of measurement, press the button Zero. 2.
ENGLISH Verifying Noise Select Tab Noise (0A) in the Performance verification interface. Standard Sample: None Measurement: 1. Remove something in the measurement channel, close the sample chamber cover, press the wavelength value, type in the wavelength of measurement, press the button Zero. 2.
ENGLISH MEASUREMENT Important Guidelines - Reagents and dilution buffers can cause cauterization and other damage to health. - Samples (nucleic acids, proteins, bacteria cultures) can be infectious and cause serious damage to health. - During sample preparation, measuring procedures and maintenance and cleaning work, observe all local laboratory safety precautions (e.g.
Page 20
ENGLISH 2. Press the icon to switch to the measurement mode. Measure absorbance value of the sample(s). Measure transmittance value of the sample(s). Measure energy value of the sample(s). 3. Press the icon to set wavelength, key in the measurement wavelength. 4.
ENGLISH Quantitation Quantitation mode is used to measure the concentration of the sample(s). 1. Main interface, press the icon to start a Quantitation application. 2. Establish Method 2.1 Quantitation interface, press the button Establish method. Page 51 Version 1, April 2024 User’s manual for UV/Vis Spectrophotometer...
Page 22
ENGLISH A=A1: Absorbance is equal to the measured absorbance value of the mea- sured wavelength 1. A=A1-m*A2: Absorbance is equal to the difference between the absorban- Measurement ce value of the measured absorbance at the wavelength 1 and the wavelen- gth 2, m is the coefficient.
Page 23
ENGLISH Measure standard sample to establish the standard curve 1- Put the reference in the measurement channel, press the button Zero to do zero. 2- Put the 1# standard sample in the measurement channel, press the button Read to measure. 3- Repeat step 2 to measure other standard samples.
Page 24
ENGLISH Input standard sample values to establish the standard curve 1- Press the item Abs and Conc to input absorbance and concentration of standard samples, press the button Next. 2.4 Establish method finished. Press the button Save to save the method, press the button Measure to accept the new method and go to the measurement interface, press the button Finish to exit.
Page 25
ENGLISH 3.2 Press the icon to select method. 3.3 Press the button Open to load measurement method stored in the internal memory/USB disk. 3.4 Press the button Measure to accept the new measurement method and back to measurement interface. 3.5 Put the reference in the measurement channel, press the icon to do zero.
ENGLISH TROUBLESHOOTING Review the information in the table below to troubleshoot operating problems. Problem Cause Solution Power on, no response Power cord connection is not reliable Improve connection Fuse burning Replace fuse Measurement uncertainty Sample is not stable Improve the sample Glass cuvettes used in UV region Use quartz cuvettes The sample concentration is too high Dilute the sample...
ENGLISH REPAIR AND MAINTENANCE Daily Maintain Check the compartment: After measurement, the cuvettes with sample solutions should be taken out of the compartment in time. Or the volatilization of the solution would make the mirror go moldy. Users must pay more attention to the corrosive sample and liquid easy to volatilize. Any solution remains in the compartment should be wipe off immediately.
Page 28
ENGLISH Reset the fuse seat: Replace the fuse seat in the power socket. Push the fuse case by using the screw- driver, and turn it clockwise, the fuse seat will be locked when released. Switch on the power: Plug the socket and switch on the power. Lamp replacement Hot! Wait 20 minutes before opening the lamp chamber after power off to...
Page 29
ENGLISH W lamp replacement: The Tungsten lamp is equipped with a blue-grey silicon coating by manufactu- rer. This coating is only a transport safety device. It can be removed with the first exchange of lamp. Pull out the defected W lamp and draw on the cotton gloves. Insert the new W lamp as deep as possible on the lamp seat.
ENGLISH WARRANTY AUXILAB S.L. warrant that this product will be free from defects in material and workmanship for a pe- riod of 2 years from date of delivery except the lamps. Lamps have a warranty of 1000 hours usage time or 6 months max.