3.
SPECIFICATIONS
3.1
FEED WATER SPECIFICATIONS
Constituent
TEA
pH
Hardness
Silica (reactive)
Silica (reactive)
TOC
Total chlorine
Fe, Mn, H
S
2
SDI 15min.
Oil & Grease
Turbidity
Oxidizing agents
Colour
TEA - MK-2E
TEA - MK-2Pharm &
PharmHT
TEA - MK-2 Mini &
MiniHT
** See section Concentrate Bleed Stream for information on recovery and feedwater
hardness.
Note: The feed water to the E-Cell Stack
rev.1.1
Units
ppm as CaCO
3
ppm CaCO
3
ppm
ppm
ppm
ppm
ppm
NTU
APHA
ppm as CaCO
* < 25 @ 15 gpm / 3.4 m
3
< 48.8 @ 7.5 gpm / 3.4 m
< 15 @ 20 gpm / 4.5 m
<26 @ 15 gpm / 3.4 m 3 /h (Nominal flow)
<55 @ 7.5 gpm / 1.7 m 3 /h (Minimum flow)
<20 @ 20 gpm / 4.5 m 3 /h (Maximum flow)
<25 @ 5 gpm / 1.14 m 3 /h (Nominal flow)
<49 @ 2.5 gpm / 0.57 m 3 /h (Minimum flow)
<15 @ 6.7 gpm / 1.52 m 3 /h (Maximum flow)
must be RO permeate or equivalent.
™
Feed
*
* As calculated by E-Calc
5 to 9
< 0.5 **
< 0.5 @ Minimum to Nominal product flow
< 0.383 @ Maximum product flow
< 0.5
< 0.05
< 0.01
<1.0
None detectable
<1.0
None detectable
<5
/h product flow (Nominal flow)
3
/h (Maximum flow)
3
/h (Minimum flow)
3
Page 3-5
Need help?
Do you have a question about the E-CELL STACK MK-2 Series and is the answer not in the manual?