Cleaning Procedure 4: Hydrochloric Acid And Sodium Chloride / Sodium Hydroxide Cleaning/Sanitizing; Cleaning Procedure 5: Halane Or Peracetic Acid Sanitizing; Cleaning Procedure 6: Hydrochloric Acid And Halane Or Peracetic Acid Sanitizing; Cleaning Procedure 7: Halane Or Peracetic Acid Sanitizing And High Ph Sanitizing - GE E-CELL STACK MK-2 Series Owner's Manual

Table of Contents

Advertisement

Cleaning Procedure 4: Hydrochloric
Acid and Sodium Chloride / Sodium
Hydroxide Cleaning/Sanitizing
This procedure provides an acid cleaning of all
chambers of the Stacks to remove inorganic
metallic scale.
Sodium chloride 5% / sodium hydroxide 1% is
useful for cleaning fouling due to organics or to
biological growth, and for sanitization. It can
also clean light silica scaling.
Steps
Cleaning Setup
Instruction 1
Instruction 6
Instruction 2
Instruction 7
Cleaning Procedure 6: Hydrochloric
Acid and Halane or Peracetic Acid
Sanitizing
Where hardness or other metallic scale is present
along with heavy biological fouling an acid
cleaning step is added to the beginning of the
procedure.
Halane, a widely used source of chlorine, is useful
for cleaning and sanitizing high-purity water
treatment equipment. Peracetic acid 0.04% /
hydrogen peroxide 0.2% performs similarly using
a different chemical species. Due to their
oxidizing power, these cleaning solutions should
be used infrequently.
Steps
Cleaning Setup
Instruction 1
Instruction 6
Instruction 5
Instruction 6
Instruction 3 or Instruction 4 (NOT BOTH)
Instruction 5
Instruction 7
rev.1.1
Cleaning Procedure 5: Halane or
Peracetic Acid Sanitizing
Halane, a widely used source of chlorine, is
useful for cleaning and sanitizing high-purity
water treatment equipment. Peracetic acid
0.04% / hydrogen peroxide 0.2% performs
similarly using a different chemical species. Due
to their oxidizing power, these cleaning solutions
should be used infrequently.
Steps
Cleaning Setup
Instruction 5
Instruction 6
Instruction 3 or Instruction 4 (NOT BOTH)
Instruction 5
Instruction 7
Cleaning Procedure 7: Halane or
Peracetic Acid Sanitizing and High pH
Sanitizing
Halane, a widely used source of chlorine, is useful
for cleaning and sanitizing high-purity water
treatment equipment. Peracetic acid 0.04% /
hydrogen peroxide 0.2% performs similarly using
a different chemical species. Due to their
oxidizing power, these cleaning solutions should
be used infrequently.
For severe biological fouling a high pH step is
added to the procedure for improved efficiency.
Steps
Cleaning Setup
Instruction 5
Instruction 6
Instruction 3 or Instruction 4 (NOT BOTH)
Instruction 2
Instruction 7
Page 11-28