When you select this parameter, the left side character field shows the message
SAMPLES. The right side character field shows the number of samples to be
taken.
Figure 3-6. Editing an Application I^rameter : Samples
■ You edit the number of samples using the Modify keys. The lower limit is 2.
The higher limit is 500. The displayed value is always the setting for the
number of samples.
■ You can press (Edit) if you have finished editing, or you can press
(Next)
or (^v)
to edit the other parameters.
AUTODUMP
When you select this parameter, the left side character field shows the message
AUTODUMP. The right side character field shows whether the automatic dump
has been enabled. The dump can be a graphic plot of the samples or a printout
of the samples. If the automatic dump is enabled the dump is made after the
Stability application has taken the last sample.
Note
Nsi
■ Other applications also use the AUTODUMP parameter. Changing
this parameter in the Logging application affects the Stability
application.
■ The Plot application sets the parameters for an automatic
plot. The Print application sets parameters for an automatic
print. These applications are described later in this chapter.
■ You should not enable automatic dumping to a printer or
plotter if there is a logging or stability application running in
the other channel. The automatic dump is given priority, and
3-10
Local Control - Menu Mode