The Standard Event Status Enable Register - HP 8153A Operating And Programming Manual

Lightwave multimeter
Table of Contents

Advertisement

Example
OUTPUT 722;">HCLS''
*ESE
standard Event Status Enable command.
Syntax
*ESE <wsp> <value>
0 < value < 266
Definition
The *ESE command sets bits in the Stai^dard Event Status
Enable register and thus enables the corresponding bits
in the Standard Event Status register. A 1 in a bit in the
enable register enables the same bit in the status register.
The value sent as an integer or a floating point number
{NRf).
The register is cleared at power-on. The *RST and *CLS
commands do not change the register.
Ihble 6-2.
The Standard Event Status Enable Register
BIT
MNEMONIC
Decimal
Value
7
Power On
128
6
Not used
0
5
Command Error
32
4
Execution Error
16
3
Device Dependent Error
8
2
Query Error
4
1
Not used
0
0
Operation Complete
1
Related Commands
*ESE?
Example
OUTPUT 722; "=kESE 21"
Remote Operation • Common Commands
6-5