POWER
FUSE REPLACEMENT
(Units Equipped with a Power Entry Module, See Figure 1)
In Units equipped with a power entry module, the power fuses are located in the power
entry module
(See Figure 1).
1.
The power cord will need to be removed to gain access to fuses. Before removing
power cord, be sure the Unit is disconnected from power source.
2.
To remove the power cord, remove the two Phillips head screws securing the cord
retainer.
3.
Detach the cord and locate the fuse door.
4.
With a small tool (i.e. small screw driver or needle nose pliers) open the fuse door
and remove the fuse carriers from the carrier socket.
5.
Replace the fuse as marked into the fuse carriers. Note:
Replace only with fuse
capacity indicated on the Unit.
Replace the fuse carriers into the socket and
~
.