Filter Replacement (500 Series) - Bair Hugger 200 Series Service Manual

Patient warming system
Table of Contents

Advertisement

1.
2.
FILTER REPLACEMENT (500 SERIES)
Change the air filter every 6 months or 500 hours.
Make sure to disconnect power to the Unit before changing the filter.
Remove the service lid by lifting it. Remove the nut from the top of the filter (See
Figure 5).
Remove
the old filter and
record the hours
indicated on the hour meter on the
Maintenance
Record at the back of this manual.
Install the new filter.
Make sure that the filter is seated firmiy to fuily engage the
interlock switch.
The Warming
Unit will not run unless the interlock switch
is
properly engaged.
Secure the filter nut and the service lid.
Plug in the Warming Unit and turn power "ON" to make sure that the filter interlock
is properly seated.
FILTER
NUT
Figure 5.
Filter Diagram (500 Series)
-10-