HOSE REPLACEMENT (200 SERIES)
1.
Remove the hose anchor screw at the blower outlet
(See Figure 1).
2.
Pull the old hose off the blower outlet flange.
3.
A hose end with pre-drilled holes must be inserted into the blanket.
4.
Attach and secure the hose anchor screw.
HOSE REPLACEMENT (500 SERIES)
1.
Grasp the hose at the base and pull free from the hose coupling flange.
2.
'
A hose end with pre-drilled holes must be inserted into the blanket.
FUSE REPLACEMENT
(200 SERIES)
(Units Not Equipped with a Power Entry Module, See Figure 1)
The
main
power
fuses
are
located
inside
the
wiring
shield
in the
hose
storage
compartment.
They are labeled with the appropriate replacement fuse capacity.
1.
Make sure to disconnect power to the Unit before replacing the fuses.
NOTE:
Replace only with fuse capacity indicated on equipment.
2.
To replace the fuses, remove the wiring shield by unscrewing the three Phillips
head screws
(See Figure 1).
3.
Unscrew the fuse holder cap, remove the old fuse, and insert the new one.
4.
Replace the fuse holder cap and the wiring shield.