RadiSense® Essntial Product Manual
RSS2010E
The RadiSense® system is a closed loop fibre system and therefore classified as a
Class 1M laser system according to EN60825-1:2014 and EN60825-2:2005. In order
to provide laser safety in case of a fibre failure or accidental disconnection of the
fibres, the RadiSense® system is provided with an Automatic Laser Shut down (ALS)
as described in the EN60825-2:2005.
Although the RadiSense® system does not emit any harmful laser light under normal circumstances, never look into any of
the fibre optic connectors.
As a safety precaution, products that use a laser can only be turned on using a laser Code. This code can be entered using
the touchscreen of the RadiCentre® system (In combination with the RadiCentre® CTR1004B or CTR1009B).
As a safety precaution, the RadiSense® series can be started by following a dedicated starting procedure as described later
in the RadiSense® product manual. (In combination with the RadiCentre® CTR1001S)
Make sure that the fibre optic cables are installed correctly before activating the system. Do not activate the system if the
fibre optic cables show any sign of damage or tampering.
Only use optic fibre cables supplied by Raditeq, or one of her official Resellers. Using alternative cables will result in (burn)
damage to the probe and fibres. the use of 3rd party cables also results in a void of warranty.
This product manual is subject to change without notice
Copyright © 2008 - 2024 Raditeq B.V.
INVISIBLE LASER RADIATION - DO NOT
EXPOSE USERS OF TELESCOPIC OPTICS
Class 1M LASER product
Max. 500 mW at 808 nm
Page 5 of 34