Download Print this page

Maintenance - Dräger PSS 5000 Series Instructions For Use

Self-contained breathing apparatus with electronic monitoring system
Hide thumbs Also See for PSS 5000 Series:

Advertisement

PSS 5000 Series with Sentinel 7000
Self-contained breathing apparatus with electronic monitoring system
Symptom
Fault
Air leak from lung demand
Ice particles on sealing
valve
elements
Air leak from quick connect
Ice particles on sealing
coupling
elements
Lung demand valve
Bypass button engaged
allowing constant air flow
into the face mask
Internal fault
High or low medium-
First-stage regulator fault
pressure
Poor sounding whistle
Whistle dirty
Whistle not functioning
Activation mechanism fault Service task
correctly
Difficulty connecting or
Dirty connector
disconnecting the medium-
pressure quick coupling
Burring of the male
coupling
Low battery indication on
Low main battery
the user interface (Fig N)
Fault code indication on
Sentinel 7000 failure
the user interface (Fig X)
Unable to switch
Low main battery
Sentinel 7000 on
Low cylinder pressure
Unknown
HUD green/yellow LED
Low HUD battery
flashing green
HUD green/yellow LED
Low main battery
flashing yellow
HUD will not log on to the
HUD out of range
Sentinel 7000
Low HUD battery
Low cylinder pressure
Battery pack sliding locks
Dirty, damaged, or faulty
not moving to the locked
components
position
6

Maintenance

6.1
Maintenance table
Service and test the breathing apparatus, including out-of-use apparatus,
in accordance with the maintenance table. Record all service details and
testing. Refer also to the instructions for use for the lung demand valve,
face mask, and other associated equipment.
Additional inspection and testing may be required in the country of use to
ensure compliance with national regulations.
Component/
Task
system
Complete
Visual inspection (see Note 1
apparatus
Section 4.6.1)
Functional testing (see Section 4.6.4)
Breathing cycle and static tests (see
Note 2)
Lung demand
Check connector for lubricant (see
valve
Note 3)
Check the male element of the quick
coupling for burring (see Step 4 in
Section 4.2)
Backup
Replace
battery
First-stage
Medium-pressure check (see Note 2)
regulator
Inspect the high-pressure O-ring (see
Note 2 and Note 4)
Cylinder
Charge cylinder to correct working
pressure
Check charged pressure (stored
cylinders only)
Check test date of cylinder (carbon
composite cylinders over 15 years old
must be retired)
Recertification
Cylinder valve Overhaul
Notes
O Dräger recommendations
1
Clean the equipment if it is dirty. If the equipment has been exposed to
contaminants, disinfect any components that come into direct and
prolonged contact with the skin.
2
These maintenance tasks may only be carried out by Dräger or trained
service personnel. Details of the tests are contained in the Technical
Manual which is issued to service personnel that have attended a
relevant Dräger maintenance course.
3
Check the O-ring on the lung demand valve. As a guide, lubricant
Remedy
should be felt on the fingers but not seen. If relubrication is required,
Press the front button
lightly apply Dow Corning
(Fig J, Item 2), allow a rush
tested and may damage the equipment).
of air to pass through the
valve, then quickly press
4
Replace the high-pressure connector O-ring if it is found to leak during
the reset button (Fig J,
functional testing or if the O-ring is visibly damaged.
Item 1) to switch off the
positive pressure.
Disconnect then reconnect
6.2
Cleaning and disinfecting
the cylinder to the
breathing apparatus (see
the quick connect coupling
instructions for use) and
CAUTION
retest.
Trapped water and ice inside the pneumatic system (such as the lung
Turn off the bypass button
demand valve) may impair the operation of the breathing apparatus.
(Fig J, Item 3)
► Prevent any liquid from entering the pneumatic system, and
thoroughly dry the breathing apparatus after cleaning and disinfecting.
Service task
Service task
NOTICE
Using cleaning and disinfecting methods not described in this section may
Clean whistle flute and
retest
damage the equipment.
► Do not exceed 140 °F (60 °C) for drying, and remove components
from the drying facility immediately when dry. Drying time in a heated
dryer must not exceed 30 minutes.
Disconnect, clean and
► Do not immerse pneumatic or electronic components in cleaning
reconnect couplings and
solutions or water.
retest
Replace the hose with the
male coupling
For information about suitable cleaning and disinfecting agents
Replace the main battery
and their specifications refer to document 9100081 on
www.draeger.com/IFU.
Service task
Refer also to the instructions for use for the lung demand valve, face mask,
Replace the main battery
and other associated equipment.
Recharge cylinder to
Use only clean lint-free cloths.
maximum working
pressure
1. Clean the breathing apparatus manually using a cloth moistened with
Service task
cleaning solution to remove excess dirt. Remove and clean the
Replace the HUD battery
following if necessary:
The removable window from the user interface screen. Clean the
Replace the main battery
window and the screen below.
The battery pack. Clean the battery and battery compartment, and
More the HUD to within
ensure that the battery contacts and locking mechanism are clean,
three feet (approximately
dry and undamaged.
900 mm) of the pressure
2. Apply disinfecting solution to all internal and external surfaces.
module
3. Rinse all components thoroughly with clean water to remove all
Replace the HUD battery
cleaning and disinfecting agents.
Recharge the cylinder to
4. Dry all components using a dry cloth, in a heated dryer or in air.
maximum working
5. Contact service personnel or Dräger if disassembly of the pneumatic
pressure
system or electronic components is required.
Attempt remedy actions in
6.2.1
Carrying harness – Thorough cleaning
this order:
1. Push down firmly on the
battery pack
NOTICE
2. Remove and clean the
Untrained personnel are not permitted to disassemble the breathing
battery pack and backplate
apparatus as they could inadvertently damage the equipment.
recess and retry
3. Replace the battery
► Thorough cleaning may only be carried out by suitably trained
pack and retry
personnel.
4. Contact Dräger
1. Remove the harness from the carrying system and clean using one of
the following methods:
a. Clean manually in a bath containing recommended cleaning or
disinfecting agents.
b. Machine wash (at 86 °F (30 °C)) using a proprietary brand
washing solution (do not use biological washing powder).
2. Rinse the harness thoroughly with clean water to remove all cleaning
and disinfecting agents.
3. Dry all components including internal parts.
6.3
Maintenance work
6.3.1
Compressed air cylinder charging
After
Every
Every
use
month
year
O
O
WARNING
Air quality for compressed air cylinders must conform to the minimum
O
O
grade requirements for Type 1 gaseous air as defined in the CGA
Commodity Specification for Air, G-7.1 (Grade D or higher quality) and,
O
where appropriate, be in accordance with: NFPA 1989 Standard on
Breathing Air Quality for Emergency Services Respiratory Protection.
O
► Ensure that the air supply meets these requirements.
O
Refer to the instructions supplied with the cylinder and the charging
apparatus for recharging a compressed air cylinder.
O
7
Transport
O
Transport the product in its original packaging.
O
8
Storage
O
8.1
Storage preparation
O
Extend the shoulder harness, waist belt, and the straps of the face
O
mask.
For storage, place the face mask in a protective bag (contact Dräger
for supply of a suitable bag).
According to national
Route pneumatic hoses in such a way that the bend radius is not too
regulations in the country of
acute and the hose is not stretched, compressed, or twisted.
use
With the system switched off, a small amount of battery power is
At the time of cylinder
consumed. If the system is not to be used for a long period, remove the
recertification
batteries (see Section 4.6.2).
8.2
Storage conditions
Store the equipment between 5 °F to 77 °F (-15 °C to +25 °C). Ensure
that the environment is dry, free from dust and dirt, and does not
subject the equipment to wear or damage due to abrasion. Do not
store the equipment in direct sunlight.
Fix the breathing apparatus securely to any raised mounting point to
prevent it from falling.
If storing the equipment in a vehicle, ensure that the breathing
apparatus is securely retained and does not interfere with the
operation of the vehicle.
®
Molykote
®
111 (other lubricants are not
Instructions for Use
9
Disposal
Dispose of the product in accordance with the applicable rules and
regulations.
9.1
Service life
All components are designed to last the lifetime of the equipment if
they are regularly inspected and maintained as described in
Section 6.1.
Retire equipment in accordance with NFPA 1852 if it can no longer be
repaired to a fully serviceable condition.
Carbon composite cylinders over 15 years old must be retired.
10
Technical data
Compressed air cylinders:
30 minutes to 60 minutes capacity.
2216 psi or 4500 psi pressure.
Composite materials.
Cylinder high-pressure connectors:
2216 psi connector to CGA 346.
4500 psi connector to CGA 347.
Quick connect coupling (2216 psi or 4500 psi).
Power supplies:
Main battery: 7.5 V.
Backup battery: 3 V.
Head-up display battery: 3 V.
RIC UAC Connector:
2216 psi or 4500 psi, male, quick coupling with relief valve.
Lung demand valve to face mask connector:
Dräger push-in connector.
EOSTI Alarms – Activation commencement range (mechanical and
electronic):
2216 psi cylinder: 819 psi to 732 psi.
4500 psi cylinder: 1665 psi to 1485 psi.
11
Special instructions
11.1
Use of an independent air supply (supplied
airline respirator (SAR) connection)
WARNING
Air quality must conform to the statutory requirements.
The time required for the wearer to escape to a safe area must be within
the remaining breathing time of the cylinder, taking into account the
remaining air content in the cylinder and the breathing rate of the wearer.
Independent air supplies must meet the following standards:
Type-1 gaseous air as defined in: CGA Commodity Specification for
Air, G-71 (grade D or higher).
NFPA 1989 Standard on Breathing Air Quality for Fire and Emergency
Services Respiratory Protection.
Air supply pressure: 87 psi to 125 psi.
Airline hose length: 5 feet to 300 feet (maximum working hose length
must not exceed 12 individual hose lengths).
Airline flow rate: 550 liters/minute.
Minimum ambient temperature of operation: -25 °F (-31.7 °C).
Sentinel 7000 alarms and warning signals will operate as normal. The user
interface display will show cylinder pressure when the cylinder valve is
open.
1. Turn on the independent air supply.
2. Connect the independent air supply coupling to the secondary supply
hose (see the UEBSS/SAR connection instructions for use) and
breathe normally.
3. Close the cylinder valve (if the whistle sounds, silence it by taking
several short deep breaths or momentarily operating the lung demand
valve bypass button (Fig J, Item 3).
4. If any air supply problems are encountered, proceed as follows:
a. Open the cylinder valve to return to breathing from the attached
cylinder.
b. Disconnect the independent air supply coupling.
c. Leave the hazardous area by the shortest and safest escape
route, if necessary.
11.2
CBRN use
The Dräger PSS 5000 Series is certified by National Institute for
Occupational Safety and Health (NIOSH), for chemical, biological,
radiological or nuclear (CBRN) use and by the Safety Equipment Institute
(SEI) to meet the requirements of NFPA 1981. Approvals are only valid
when the apparatus is used with compressed-air cylinders approved by
NIOSH. Equipment configurations for CBRN use are detailed in a CBRN
Approval Label (see CBRN Approval Label 3367113).
Dräger recommend that a quantitative fit test (QNFT) be performed on the
face mask before use in a CBRN environment. The fit test must be
conducted strictly in accordance with the requirements outlined in the
OSHA Respiratory Protection Standard 29 CFR, Section 1910.134.
11.3
Cautions and limitations
CAUTIONS AND LIMITATIONS
D – Air-line respirators can be used only when the respirators are supplied
with respirable air meeting the requirements of CGA G – 7.1, Grade D
or higher quality.
E – Use only the pressure ranges and hose lengths specified in the
USER'S INSTRUCTIONS.
I – Contains electrical parts that may cause an ignition in flammable or
explosive atmospheres.
J – Failure to properly use and maintain this product could result in injury
or death.
M – All approved respirators shall be selected, fitted, used, and
maintained in accordance with MSHA, OSHA, and other applicable
regulations.
3367112 (A3-D-P) Page 4 of 5

Advertisement

loading