Chapter 5 Maintenance; Keeping Maintenance Records; Clean The Detector - Emerson Rosemount 975 Manual

Flame detectors
Hide thumbs Also See for Rosemount 975:
Table of Contents

Advertisement

Manual
00809-1400-4975
5 
Maintenance
5.1 

Keeping maintenance records

Record all maintenance operations performed on a detector in accordance with site
guidance and requirements.
5.2 

Clean the detector

Procedure
1. Disconnect power from the flame detector.
2. Wipe the detector housing with clean water and a damp cloth.
3. Identify where dust, dirt, or moisture accumulates on the detector window.
Rosemount 975
NOTICE
Do not use a brush or sharp tools.
a) Clean with a soft optical cloth.
b) Rinse with clean water.
Maintenance
September 2023
29