5.4.2.2. Wi-Fi Access Point

In this mode, analyzer is configured as a Wi-Fi Access Point. Turn the knob on the analyzer to Wi-Fi
access point option. On the laptop/tablet, analyzer can be found in available wireless networks. For
example, if analyzer's serial number is 12345, SSID visible in available networks would be
Gasmet_SN12345. Default password is password64. Connect to this Wi-Fi access point from
laptop/tablet. In the Configuration dialog, select TCP/IP mode and enter IP address and port number.
Default IP address is 192.168.64.64 and port is 8080.

5.4.2.3. Bluetooth

If laptop/tablet and analyzer are already paired via Bluetooth, Mode can be selected to Bluetooth in
Configuration dialog and analyzer's Bluetooth address can be entered. This address can be found
from Control Panel's 'Device and Printers' option. Right click on paired Gasmet device and select
'Properties'. Navigate to the tab 'Bluetooth' and write down the 'Unique identifier', twelve
hexadecimal digits. This is the analyzer's Bluetooth address based on which Calcmet will find the
correct analyzer. Baud rate to be used for this option is 115200.
If they are not already paired, instructions about how to pair are explained in Calcmet Easy user
manual. As pairing needs exchange of randomly generated six digits number, that is possible using
Analyzer Setup options only.
5.4.3.

Remote use

Calcmet Easy can be utilized to operate the analyzer remotely. For this purpose, Wi-Fi station or
Ethernet options can be used. Make sure both devices are in the same network.

5.5. Calcmet setup

This chapter lists the most important settings in Calcmet Easy. For further information, see Calcmet
Easy manual.

5.5.1. Autosaving

Select Autosave in the menu. Select which files are saved and where. It is recommended to save
both the results and the sample spectra. Sample spectra can be re-analyzed later if needed as the
original raw data is stored, whereas result files only contain the results of the initial analysis. Make
sure that data is being saved before starting the measurements.

5.5.2. Selecting measuring times

Select Measuring Times to configure flush time, pump time, sampling time and measuring interval
for continuous measurements.

5.5.3. Selecting the application

Select and load the application by clicking Select Application in the menu.

5.5.4. Selecting the measured component and their settings

If necessary, edit the analysis settings by selecting Edit Application in the menu.