Spare parts may be new or equivalent to new. Gasmet Technologies Oy and its subsidiaries agree to either replace or repair free of charge (Ex Works Helsinki, Incoterms 2010) any such defective product or part that is returned to its repair facility within one (1) year of the delivery date.
- Ensure that a sample tube is connected to SAMPLE IN connector before starting the sample pump. Otherwise the valve in the sample inlet will not open. - The GT5000 Terra Instruction and Operating Manual should be read and understood before operating your unit.
Appendix E: CERTIFICATION ................... 46 Figures Figure 1. Attaching the backpack harness to GT5000 Terra. 1. Attach the connective metal pieces to the loops on the back of the analyzer. 2. Place the harness on top of the analyzer. Align it so that the loops on the sides of the metal pieces go through the holes on the harness.
Table 3. List of files after Calcmet setup..................16 Table 4. Communication settings....................17 Table 5. Maintenance intervals for GT5000 Terra................ 27 Table 6. Power switch in Off mode....................33 Table 7. Power switch in Cable mode..................34 Table 8.
Calcmet Expert software complement this manual and are referred to in matters considering the Calcmet software. Before starting to operate the GT5000 Terra gas analyzer, read this Instruction & Operating Manual carefully and thoroughly. Improper use of the analyzer may damage the equipment.
> GT5000 Terra FTIR gas analyzer > AC/DC PSU for GT5000 Terra > USB cable for GT5000 Terra > Gasmet USB flash drive including software, manuals and other relevant documentation The options available are: > Gasmet tablet > Gasmet probe with one PTFE filter, flexible Tygon tube and connectors >...
When operating the analyzer, place the analyzer in vertical position or in horizontal position front panel facing the user. Carry the analyzer from the handle or with the optional Gasmet backpack harness. Do not use any carrying equipment that is not approved by Gasmet. See below instructions...
> +10 - +35 ↓C storage temperature > 5 - 95 % relative humidity, non-condensing > Gasmet battery should be stored at 15 % to 50 % state of charge with an ambient temperature preferably not exceeding +30 °C. For short term storage, e.g. during shipping, the following conditions can be tolerated: >...
4.4.2. Ventilation Requirements Care should be taken in order to ensure a free convection airflow around the GT5000 Terra and avoid fan obstruction when in long term operation with AC/DC PSU. Do not block fans air circulation. This would decrease life expectancy of most of the internal components.
Follow the direction of the arrow located on top of the battery in order to insert the battery in the correct way into the GT5000 Terra analyzer. Ensure that the battery is fully inserted. If using the analyzer in wet conditions, ensure that the battery compartment lid is properly closed.
Visual C++2015 Redistributable Package Table 1. C++ redistributable packages required for Calcmet installation. It is recommended that the program would be installed to its default location in C:\Gasmet Technologies\Calcmet\ directory. The software works on other directories too, but the demo library settings will not work, as they need the files to be located in these specific folders.
.CLIB file under the C:/ from Gasmet USB. If the reference spectrum files (*.REFX) and library files (*.CLIB) were copied from the Gasmet USB, it may be that the files are write protected. To later make any changes to settings, it will be necessary to deactivate write protection from these files.
Gasmet GT5000 Terra Instruction and Operating Manual 5.4. Startup and connecting to PC/tablet Connecting to analyzer can be divided into two parts. One relates to configuration of laptop/tablet and other relates to configuration of the analyzer. In this section, only settings for the laptop/tablet are discussed.
Configuration dialog and analyzer’s Bluetooth address can be entered. This address can be found from Control Panel’s ‘Device and Printers’ option. Right click on paired Gasmet device and select ‘Properties’. Navigate to the tab ’Bluetooth’ and write down the ’Unique identifier’, twelve hexadecimal digits.
Gasmet GT5000 Terra Instruction and Operating Manual 5.6. Hardware status Before making the background measurement, ensure that the analyzer has been powered up for at least 30 minutes. At the end of this warm-up time, measure the hardware status by selecting Hardware.
You can purge continuously or fill the interferometer with zero gas and measure in the field. GT5000 Terra enclosure is tight and will hold majority of the purge gas in the interferometer for hours. If you are performing extremely high accuracy measurements, purge for longer time or increase the zero gas flow to 2.0 l/min.
Gasmet GT5000 Terra Instruction and Operating Manual 5.8. Background measurement Perform a background measurement with nitrogen every day before measurements. Good background is required to obtain accurate analysis results. Connect nitrogen bottle with a regulator and flow controller to the zero gas in in the front panel. 5.0 purity nitrogen should be used for the background measurement.
Gasmet GT5000 Terra Instruction and Operating Manual Figure 4. Results view shows the results of the analysis. Residual-column shows the difference between the sample spectrum and the calculated model spectrum in absorption units. If the residual value is too high, a warning will appear. High residual values can indicate the presence of an unknown component.
Gasmet GT5000 Terra Instruction and Operating Manual Figure 6. In results view components can be hidden or displayed by right-clicking the component. 5.9.2.2. Spectrum View In the spectrum view, you can see the sample spectrum. This is the default view displayed when you open Calcmet Easy.
Gasmet GT5000 Terra Instruction and Operating Manual 5.9.2.3. Residual View Figure 9. Residual view shows the residual spectrum. Zooming buttons are located on the right. Residual view shows the difference between the sample spectrum and the calculated model spectrum as a spectrum.
Gasmet GT5000 Terra Instruction and Operating Manual Figure 10. Trend view shows the concentration of selected components as a function of time. Figure 11. Components can be selected to be shown in the trend view by clicking the number of the component.
> Intensive shocks may damage the analyzer. Be always very careful when you move or transport the instrument. > Use a soft cloth to wipe off the dirt on the case. Do not use strong detergents or acetone. > Never replace Gasmet Power supply cord by an inadequately rated cord.
Gasmet GT5000 Terra Instruction and Operating Manual 6.2. Maintaining the analyzer 6.2.1. Maintenance plan To keep your analyzer in optimal condition, maintenance is recommended at certain intervals. Recommended maintenance and intervals are defined in Table 5. After maintenance, the analyzer is visually inspected and correct operation is ensured.
> Operational test. 7. GT5000 Terra technical description 7.1. Analyzer Enclosure GT5000 Terra specially designed enclosure is made of ABS/PC, material used for hard-plastic parts. In order to comply with EMC (Electromagnetic Compatibility), the enclosure receives a thin layer of copper coating.
If the battery voltage is too low (below +9 V) and the GT5000 Terra is switched ON only under battery mode, the analyzer will remain off in order to prevent voltage depletion and damage to the battery. For best performance, keep the battery within 15 - 50 % of full charge for long-term battery storage.
USAGE INSTRUCTIONS and Appendix C: RECOMMENDATIONS TO THE END-USERS OF BATTERIES. 7.2.2. AC/DC PSU A 120-watt AC/DC PSU is used to power the GT5000 Terra and to charge the battery. DC power supply input is located on the side of the analyzer, behind the battery hatch. The required mains input is from 100 VAC up to 240 VAC, 50 Hz- 60 Hz.
Figure 15. GT5000 Terra front panel. 7.3.1. Gas Fittings All gas connectors in the GT5000 Terra analyzer are grouped together on the front panel. The quick- connect fittings accept 6 mm OD PTFE or PFA tubing and the Gasmet 6 mm tube adapter, which allows the use of more flexible tubing such as Tygon.
Gasmet USB or Ethernet cable is connected to the communication port. 7.3.3. Power Switch and LEDs GT5000 Terra is started up using the rotary switch located in the top right end corner of the front panel. Around the rotary switch, five LEDs display the status of charge of the battery.
Gasmet GT5000 Terra Instruction and Operating Manual 7.3.3.1. Power Switch in Off Mode Table 6 below sums up the different options possible when power switch is in Off mode. OPTION CURRENT ANALYZER CHARGE OF BATTERY Battery not No current OFF –...
Gasmet GT5000 Terra Instruction and Operating Manual 7.3.3.2. Power Switch in Cable Mode Table 7 below sums up the different options possible when power switch is in Cable mode. CURRENT ANALYZER CHARGE OF BATTERY Battery not connected No current OFF – Nothing...
Gasmet GT5000 Terra Instruction and Operating Manual 7.3.3.3. Power Switch in WiFi Access Point Mode Table 8 below sums up the different options possible when power switch is in WiFi Access point mode CURRENT ANALYZER CHARGE OF BATTERY Battery not connected No current OFF –...
Gasmet GT5000 Terra Instruction and Operating Manual 7.3.3.4. Power Switch in Bluetooth Mode Table 9 below sums up the different options possible when power switch is in Bluetooth mode. OPTION CURRENT ANALYZER CHARGE OF BATTERY Battery not connected No current OFF –...
Gasmet GT5000 Terra Instruction and Operating Manual 7.3.3.5. Power Switch in WiFi Station Mode Table 10 below sums up the different options possible when power switch is in WiFi station mode CURRENT ANALYZER CHARGE OF BATTERY Battery not connected No current OFF –...
Table 18. Charger communication error. 7.3.4. On-board Sample Pump The GT5000 Terra model is equipped with an on-board sample pump, which has a neoprene diaphragm membrane. Pump power supply is +12 VDC, maximum pressure is 1.4 bar and flow is 2 l/min.
Gasmet GT5000 Terra Instruction and Operating Manual GT5000 Terra specifically designed light probe combines an embedded gas filter. Filter cleanliness is important to insure correct measurements. Filters provided by Gasmet Technologies Oy are the only one usable. Sample gas should be at ambient temperature and avoid sudden changes in ambient gas temperature in order to prevent condensation in sample cell.
8. If you cannot charge the battery, it may be overcharged. Discharge the battery and check again. 9. If the battery has been short-circuited, charge it to reset the electronics with the Gasmet charger. 10. If the battery has been short-circuited, status of charge can be wrong. Charge the battery for ½...
Gasmet GT5000 Terra Instruction and Operating Manual Appendix C: RECOMMENDATIONS TO THE END-USERS OF BATTERIES The following represents a typical, but non-exhaustive, list of good advice to be provided by the manufacturer of secondary cells and batteries to equipment manufacturers and battery assemblers.
Page 44
Gasmet GT5000 Terra Instruction and Operating Manual normal room temperature. r) Retain the original cell and battery literature for future reference. s) When disposing of secondary cells or batteries, keep cells or batteries of different electrochemical systems separate from each other.
GT5000 Terra user should be aware that the apparatus contains a potentially harmful laser device. GT5000 Terra fulfils the requirements for Laser Class 1 according to the standard EN 60825-2007 (IEC 60825-2007). The laser device, included into the GT5000 Terra, is fully enclosed and only trained service personnel should open the enclosure in case of service of the analyzer.
Gasmet GT5000 Terra Instruction and Operating Manual Appendix E: CERTIFICATION This device contains licence-exempt transmitter(s)/receiver(s) that comply with Innovation, Science and Economic Development Canada’s licence-exempt RSS(s). Operation is subject to the following two conditions: This device may not cause interference.