9 mOPC Errors

Each mini-OPC error is assigned a bit in the ERROR number. Convert the
ERROR number to binary to check the bits. If an error is occurring, the bit
for that error will be 1 instead of 0.
Bit Number
ERROR
0
Bad Stored Value
1
PMT Self Cal ERROR
2
Bad Thermistor ERR
3
Digital Pot ERROR
4
RH Sensor ERROR
5
Flow ERROR
6
Max Laser Power (laser too warm)
7
SD Card ERROR
8
HG Pulse Calib ERR
9
LG Pulse Calib ERR
10
Clock Battery ERROR
11
Clock Chip R/W ERR
12
RTC/EEPROM FAILED
13
EEPROM Address ERR
15
Clock Crystal ERROR
An example error number, converted from HEX to Binary might be: 000000010001,
bit number 1 and 5 are both 1, indicating error 1: Bad Stored Value and er-
ror 5: RH Sensor ERR.