1.2 Shock during shipping If the shock watch installed on the shipping case has tripped, contact BRECHTEL. Oper- ate the unit to confirm proper operation before using for actual measurements.
Contact BRECHTEL 3 Contact BRECHTEL 3.1 Getting Help Contact BRECHTEL for any questions or assistance with this product. Include the model and serial numbers when contacting Brechtel. 1789 Addison Way Hayward, California 94544 Telephone: (510) 732-9723 Fax: (510) 732-9153 Email: sales@brechtel.com...
Unboxing 4 Unboxing Each BRECHTEL mini-OPC is inspected and tested in-house at BRECHTEL to ensure out of the box operation upon delivery. Prior to opening, inspect the packaging container and ShockWatch impact indicators. Carefully open the package and inspect the instrument and any accessories for broken parts, scratches, dents or other signs of damage incurred during shipping.
Do NOT change the values of the sample and sheath flow rates from 0.06 and 0.6 lpm, respectively. Doing so invalidates the size calibration of the mini-OPC. 83-00032-01 mini-OPC 9405 Page 8 of 39...
Page 10
Particles take up water under conditions of high humidity, which is one reason the mini- OPC sample flow should be dried before entering the unit to remove particle-bound wa- ter that will influence particle size. Contact Brechtel if you require a sample flow drying solution.
To improve the accuracy of the measured concentrations, the fol- lowing equation may be used to correct the measured concentrations for coincidence: 83-00032-01 mini-OPC 9405 Page 10 of 39...
Note: Saved concentration data have NOT been corrected for coinci- dence. 5.3 Component Technical Descriptions The BRECHTEL mini-OPC system includes three separate components. Component 1 mini-OPC main assembly (body, detector board, main board, pump all integrated to ’U’ bracket mount) Component 2 A front panel display connected by ribbon cable to the main board.
Additional 12 VDC power supply Duplicate supply to operate mini-OPC in for ground operation of any two locations ACCESS module ACC-PC Computer loaded with UAV Easily operate mini-OPC in multiple Reader Software locations. 83-00032-01 mini-OPC 9405 Page 12 of 39...
3.7 x 4.7 x 6.0 in/9.4 x 11.9 x 15.2 cm Weight 2.2 lbs/1.0 kg Supply voltage 12 VDC (9-16 VDC) Power usage 6-9 watts (without/with display) Mounting Hard mounting points in metal body bracket 83-00032-01 mini-OPC 9405 Page 13 of 39...
Table 3: Assembly Parts Description Base Unit Front Panel Display Display Ribbon Cable Ship Kit Power supply Figure 4: View of mini-OPC assembly. 83-00032-01 mini-OPC 9405 Page 14 of 39...
Ensure the power plug is fully seated in the connector before connecting the instrument to a computer. 6.5 Communication Connections Connect the various cables provided with the mini-OPC as described in the table below. 83-00032-01 mini-OPC 9405 Page 15 of 39...
USB Port on the mini-OPC main board as shown in figure 4. The UAV Reader software may be purchased as an accessory to the mini-OPC installed on a ruggedi- zed tablet, contact BRECHTEL for information. Connect the USB cable prior to running the software. Do not disconnect the USB cable while the software is running.
Bolts should not extend into the cover so much that they would interfere with the main control board connections. Figure 5: mini-OPC as- sembly overall dimensions (inches and millimeters in parentheses). 83-00032-01 mini-OPC 9405 Page 17 of 39...
Page 19
6.6 Mounting Figure 6: Front panel dis- play mounting bracket de- tailed dimensions (inches and millimeters in paren- theses). 83-00032-01 mini-OPC 9405 Page 18 of 39...
To protect the detector, good ESD practices are critical to main- taining and operating your device. • Be sure to always ground yourself before handling the unit • Be sure to always connect the unit to earth ground 83-00032-01 mini-OPC 9405 Page 19 of 39...
The only user-settable item in this menu is ’SD Save’, all other items are set by the mini- OPC automatically. A quick push of the data knob will select the SD Save value, rotate the knob to change the value. 83-00032-01 mini-OPC 9405 Page 20 of 39...
0-127 (startup setting for pmt_base) DIAGNOSTIC Menu pmt_base_pot_adj: (current PMT base pot setting) pmt_base_rd: (current PMT base reading) ERRORS Menu Push data knob to select Error Number. Scroll through errors. Currently occurring er- rors will have ! 83-00032-01 mini-OPC 9405 Page 23 of 39...
”SS” is the second. As new files are created on a given day they are named with the corresponding date and time at which the acquisition started. Turning SD Card save to ”ON” will save the same format data files described below to the SD Card. 83-00032-01 mini-OPC 9405 Page 24 of 39...
#pmt_gain_pot=37 (PMT gain setting) #lsr_brt_pot=110 (Laser brightness pot setting) #lsr_gain_pot=127 (Laser gain pot setting) # (blank line) #cal_hg1=170,460 (high gain size cal, point 1 of 20) #cal_hg2=180,568 #cal_hg3=190,760 #cal_hg4=200,1032 #cal_hg5=210,1392 #cal_hg6=220,1792 #cal_hg7=250,4096 #cal_hg8=260,4704 #cal_hg9=280,5760 #cal_hg10=310,6784 83-00032-01 mini-OPC 9405 Page 25 of 39...
Page 27
Below the header data in each data file, the numeric data is stored in columns as follows: 1. YY/MM/DD: Year, Month, Day 2. HR:MN:SC: Hour, Minute, Second 3. opc_cntl: opc control (0=off, 1=on) 4. sample_sp: sample flow setpoint (lpm) 83-00032-01 mini-OPC 9405 Page 26 of 39...
The UAV Reader software has been designed to allow multiple instruments to be con- nected simultaneously so data can be logged and instrument performance monitored all using one user interface. Each instrument has a ”main” tab (display) in the software as 83-00032-01 mini-OPC 9405 Page 27 of 39...
Page 29
There are four different data displays of the number size distribution data on the OPC main tab in Fig. 7. The rightmost tab displays the 84 bins of 1Hz raw count data versus bin number. This is a useful display for checking size calibrations. The total number 83-00032-01 mini-OPC 9405 Page 28 of 39...
Page 30
The calibration coefficient values for the sheath and sample LFEs (quadratic), pressure (linear), and the PMT baseline are shown on the details tab. DO NOT change the calibration coefficients unless you have re-calibrated a 83-00032-01 mini-OPC 9405 Page 29 of 39...
Page 31
Each time the reader software initiates a new mini-OPC run, the mini- OPC real time clock is synchronized with the clock on the computer run- ning the reader software. Be sure the computer clock is set to the desired time. 83-00032-01 mini-OPC 9405 Page 30 of 39...
Page 32
- sets sheath flow setpoint (lpm) o_clk_sec - set clock seconds o_clk_min - set clock minutes o_clkhour - set clock hours o_clk_day - set clock day o_clk_mon - set clock month o_clkyear - set clock year 83-00032-01 mini-OPC 9405 Page 31 of 39...
RS232 port (e.g. NOT using the reader software): ”read” total_conc=0.0 sample_flw=0.00 (lpm) sheath_flw=0.00 (lpm) sheath_temp=23.9 (deg C) 83-00032-01 mini-OPC 9405 Page 32 of 39...
Remove the nozzle from the unit and flush with clean DI water and IPA. Sonicate for 2 hours submerged in water. Blow dry with filtered air after sonication and re-install. 83-00032-01 mini-OPC 9405 Page 37 of 39...
An example error number, converted from HEX to Binary might be: 000000010001, bit number 1 and 5 are both 1, indicating error 1: Bad Stored Value and er- ror 5: RH Sensor ERR. 83-00032-01 mini-OPC 9405 Page 38 of 39...
Laser failure Replace laser board Sample flow incorrect Calibrate LFE Sample nozzle clogged Clean nozzle Sizing incorrect Calibration shift Recalibrate Laser shift Replace laser board Sample flow out of range Check LFE calibration 83-00032-01 mini-OPC 9405 Page 39 of 39...
Page 41
Flash Memory, SD Card, 4 GB, Micro and Adapter 903100031A Memory Card Reader, USB 96-00004-01A IM2450 - foam filled (Pick n Pluck) - black case 96-00007-02A Shock Label, 7/8” x 2”, 25 G Sensitivity, Resettable POWER-OPTION POWER CORD (OPTION by Country/Voltage) 83-00032-01 mini-OPC 9405 Page 40 of 39...