Common operations
Following operations apply to CuroCell® iA Automatic and CuroCell® iA Manual regardless of which
mattress that is used.
3.1 CPR (Cardiopulmonary resuscitation)
In case of an emergency where CPR
(Cardiopulmonary resuscitation) is necessary,
remove the connection (marked 'CPR') from the
control unit and leave the lid open to empty the
mattress of air quickly.
3.2 Restart
If a restart is required, set the On/Off switch on
the side of the control unit to 0 (Off). Wait for
approx. 10 seconds and restart the control unit.
3.3 Maximum pressure notification
When the function Maximum pressure
has been used for a long time, the
Maximum pressure diode will blink.
If the use is intentional, ignore the notification.
3.4 Power failure
In the event of a power failure, the valves will
open automatically and even out the air pressure
in the mattress. The mattress will retain air for at
least 12 hours. Perform a hand check to make
sure the pressure of the mattress is not too hard
or too soft. See 4.6 for CuroCell® iA Automatic
and 5.7 for CuroCell® iA Manual for more
information about hand check.