Reuse and cleaning

The product is reusable. Before reusing, it is important to follow the instructions below for cleaning,
disinfection, and reconditioning. Disinfection is recommended between patients according to the
instructions below.
Always follow local instructions and the instructions for use of the cleaning and disinfecting agent.
Consult your hygiene manager or Care of Sweden for help and instructions in case of uncertainty.
7.1

Cleaning and disinfection

CONTROL UNIT
Wipe the control unit with a damp cloth and mild detergent. Primarily use solvent-free soap
with a neutral pH value. If necessary, a disinfectant and/or cleaning agent can be used such
as: alcohol with or without surfactants or oxidizing solutions such as: chlorine and/or
hydrogen peroxide, concentration 1000 ppm/0,1%. In exceptional cases, a maximum concentration of
10,000 ppm/1% can be used.
If another agent is used, choose one that does not harm the exterior of the control unit.
INNER COVER AND MATTRESS COVER
Wipe off
Primarily use solvent-free soap with a neutral pH value. For a daily basis cleaning, a
disinfectant and/or cleaning agent can be used such as: alcohol with/without surfactants or
oxidizing solutions such as: chlorine and/or hydrogen peroxide, concentration 1000ppm/0,1%.
In exceptional cases, a concentration of a maximum of 10,000ppm/1% can be used, then consider that
high concentrations can shorten the life of the coating.
Mechanical cleaning
Covers consisting of several parts must be separated before washing.
7.2

Reconditioning

CONTROL UNIT
Clean the control unit according to section 7.1 Cleaning and disinfection – Control unit.
MATTRESS
Disconnect the tube connector from the control unit and remove the air from the mattress.
Cleaning of mattress
1. Clean all external surfaces of the mattress according
to section 7.1 Cleaning and disinfection - Inner cover
and Mattress cover - Wipe off. Ensure that all areas
are free of dirt residues.
2. If the mattress is heavily soiled, it is recommended
that the mattress is taken apart and cleaned, follow
the instructions below according to points 3-5.
3. Remove the covers.
4. Wipe off the cells, tubing and the CPR module with
a cleaning agent according to local instructions and
the instructions for use of the cleaning and disinfecting agent.
5. When all parts are dry, assemble the mattress. If cells have become detached from the tubes, these
must be put back according to drawing in section 6.3.
Note!
Check the hygiene cover, cells and
hoses at each cleaning. If the hygiene
cover is damaged, it must be replaced.
Also check the control unit, tube
connections and power cord when
cleaning. Damaged parts must be
replaced or repaired.