Service and maintenance

The control unit should be regularly serviced and inspected to maintain functionality and
performance. Only use spare parts approved by Care of Sweden.
9.1

Service and maintenance chart

Service in the column to the right must only be performed by an authorized service technician.
Control unit
Visual inspection of exterior
Cleaning of exterior
Visual inspection of power
cable/supply
Test function on control
panel
Visual inspection of the
connecting tubes (marked
CPR), they shall be
positioned correctly and not
leaking
Control air filter and replace
if needed
Mattress and cover
Visual inspection of cover,
no damages
Test function of zipper
Visual inspection of
connecting tubes
Other
Function test (start-up)
Full service
Change valve module
Before every use (or
every 2 weeks if used
continuously)
X
X
X
24
After each
After 12
use
months
(between
users)
X
X
X
X
X
X
X
X
After 5 years
X
X