Hand Check (Function Control); Cpr (Cardiopulmonary Resuscitation); Restart; Power Failure - Care of Sweden CuroCell iA Automatic Manual

Table of Contents

Advertisement

5.7

Hand check (function control)

Hand check is performed to ensure that the
mattress system works properly and to ensure
that the weight setting is correct. Hand check
should be performed regularly; for CuroCell® iA
Manual, we recommend once every eight hours
as well as after installation of the system.
For manual systems hand check should also be
performed when moving from lying to seating
position in bed, in case of positioning changes
and when changing the comfort setting (and/or
the weight setting).
Note!
• Make sure that the mattress is filled, and
that the system has adapted to the
patient before performing a hand check.
• How to perform a hand check depends
on which mattress is used – read the
label of the mattress carefully to know
which mattress you have.
When using an overlay mattress system
(CuroCell® Ci10 or CuroCell® Ci10 PRO):
1a. Open the cover and identify a cell with less
air at the patient's sacrum. Insert a hand, with
the palm facing up, between the overlay
mattress and the underlying mattress. The hand
is inserted beneath the patient's sacrum (center
of mattress).
When using a full replacement mattress
(CuroCell® Ci17, CuroCell® Ci20, CuroCell® Ci17
PRO or CuroCell® Ci20 PRO)
1b. Open the cover and identify a cell with less
air at the patient's sacrum. Insert a hand, with
the palm facing up, between the top cells and
the underlying safety mattress. The hand is
inserted beneath the patient's sacrum (center of
mattress).
2. Ensure there is a gap between the patient and
the underlying mattress so that the patient does
not 'bottom out'.
3. If you can feel the patient's sacrum resting in
the palm of your hand, the gap is too small. See
section 10 'Troubleshooting'.

5.8 CPR (Cardiopulmonary resuscitation)

See 3.1
5.9

Restart

See 3.2
5.10

Power failure

See 3.4
5.11

Notifications

Different notifications exist based on
how serious the warning is. With a
malfunction or an error, a notification
will be given by a flashing warning
triangle. To mute the warning signal,
press the mute button.
will turn off and a notification code is shown on
the four different cycle time diodes (10, 15, 20,
25). To read the cycle time during the error
notification, unlock the control panel.
5.12

Maximum pressure notification

See 3.3
15
CuroCell® iA Manual