6 Replacing the Battery or the Fuses
A If nothing is displayed or if the symbol
insert a new 9 V battery .
B If it is not possible to measure currents, check the internal fuses and
have them replaced by skilled personnel .
You will find the battery compartment and the fuse holders under the
housing cover on the rear side of the meter .
1) Fold up the tilt stand, release the two screws of the housing cover,
then remove the cover .
2) To replace the battery, remove the old battery, connect the battery
contacts to the new battery and insert it into the compartment .
Always replace a defective fuse by one of the same type (see Speci-
fications) .
3) Fasten the screws of the cover again before setting the meter into
operation .
7 Specifications
Display: . . . . . . . . . . . . . 3 ½ digits, 21 mm LCD
Measuring rate: . . . . . . . 2 measurements per second
Measuring category: . . . CAT III 250 V
Voltage measurement: . . ⎓ /~ 250 V max .,
Fuses
up to 200 mA: . . . . . . F 200 mAL / 250 V
200 mA to 10 A: . . . . . F 10 AL / 250 V
18
WARNING: Before opening the housing, remove the test
tips from the measuring points and the test leads from the
jacks on the meter, otherwise you will risk an electric shock .
Never operate the meter when it is open .
Never put batteries in the household waste . Always dispose
of the batteries in accordance with local regulations .
200 mV range: overload protection ⎓ /~ 200 V
(= low battery) is displayed,