Iodine
Preliminary examination:
Please examine this product carefully. Make sure that the instrument is not damaged. If
any damage occurred during shipment, please notify your Dealer.
Each HI 718 meter is supplied complete with:
• Two Sample Cuvettes and Caps
• Six packets of Iodine Reagent
• 1 x 1.5V AAA Battery
• Instruction Manual
Range
Resolution
Accuracy
Light Source
Light Detector
Method
Environment
Battery Type
Auto-Shut off
Dimensions
Weight
w w w . h a n n a i n s t . c o m
For more details about spare parts and accessories
see "Accessories".
Technical specifications:
0.0 to 12.5 ppm
0.1 ppm
±0.1 ppm ±5% of reading @ 25 °C / 77 °F
Light Emitting Diode @ 525 nm
Silicon Photocell
Adaptation of the Standard Methods for the Examination
of Water and Wastewater, 18
th
edition, DPD method.
The reaction between iodine and the reagent causes a
pink tint in the sample.
0 to 50 °C (32 to 122 °F);
max 95% RH non-condensing
1 x 1.5V AAA
After 10 minutes of non-use
81.5 x 61 x 37.5 mm (3.2 x 2.4 x 1.5")
64 g (2.25 oz.)
Functional description:
1. Dust cover.
2. Cuvette with cap.
3. Cuvette holder.
4. Liquid Crystal Display.
5. Button.
Errors and warnings:
Light High: There is too much light to perform a measurement.
Please check the preparation of the zero cuvette.
Light Low: There is not enough light to perform a measurement.
Please check the preparation of the zero cuvette.
Inverted Cuvettes: The sample and the zero cuvette are inverted.
Under Range: A blinking "0.0" indicates that the sample absorbs
less light than the zero reference. Check the procedure and make sure
you use the same cuvette for reference (zero) and measurement.
Over Range: A flashing value of the maximum concentration indicates
an over range condition. The concentration of the sample is beyond
the programmed range: dilute the sample and re-run the test.
Battery Low: The battery must be replaced soon.
Dead Battery: This indicates that the battery is dead and must be
replaced. Once this indication is displayed, normal operation of the
instrument will be interrupted. Change the battery and restart the
meter.
www.ponpe.com

Advertisement

loading
Need help?

Need help?

Do you have a question about the Iodine HI 718 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Hanna Instruments Iodine HI 718

  • Page 1 Dear Customer, Functional description: Thank you for choosing a Hanna Instruments Product. Please read this instruction manual carefully before using the instrument. If you need HI 718 additional technical information, do not hesitate to e-mail us at tech@hannainst.com or view our worldwide contact list at www.hannainst.com.
  • Page 2 Add the content of one • Turn the instrument off by holding the button until the meter shuts off. Hanna Instruments reserves the right to modify the design, construction or packet of HI 718-25 reagent. Replace the •...