Figure 9.1
Malfunction (Failure):
Due to a malfunction in the transmitter, a fault signal is output.
This status message is always active. It cannot be deactivated by the user.
Function check:
The transmitter is being worked on, the measured value is temporarily invalid (for
example during simulation).
This status message is inactive by default.
Out of specification:
The measured value is unreliable because an transmitter specification was exceeded
(e.g. electronics temperature).
This status message is inactive by default.
Maintenance required:
Due to external influences, the transmitter function is limited. The measurement is
affected, but the measured value is still valid. Plan in maintenance for the transmitter
because a failure is expected in the near future (e.g. due to buildup).
This status message is inactive by default.
Failure
Code
Text message
F013
no measured value
available
F017
Adjustment span too
small
F025
Error in the
linearization table
F036
No operable software
F040
Error in the
electronics
F080
General software
error
SITRANS LR120 Modbus
Operating Instructions, 10/2022, 62709-76-01, A5E49829030
Maintenance required - blue
Pictographs of the status messages
Cause
No measured value in the switch-on
phase or during operation
Transmitter tilted
Adjustment not within specification
Index markers are not continuously
rising, for example illogical value pairs
Checksum error if software update
failed or aborted
Limit value exceeded in signal
processing
Hardware error
General software error
Diagnostics and troubleshooting
9.3 Status messages according to NE 107
Rectification
Check or correct installation and/or
parameter settings
Clean the antenna system
Change adjustment according to the
limit values (difference between min.
and max. ≥ 10 mm)
Check linearization table
Delete table/Create new
Repeat software update
Send transmitter for repair
Restart transmitter
Send transmitter for repair
Restart transmitter
33