(4) Cautions for storage

To avoid failures or damages, do not store this product in the places where:
・ exposed to water
・ susceptible to adverse effects due to air pressure, temperature, humidity, ventilation,
Cautions
・ exposed to inclination, vibration or shock (including during transportation)
・ exposed to chemicals (chemicals' storage area) or gas
・ exposed to direct sunlight or high temperature in a car, etc.
9.

Operation

(1) Check whether the wire connection is correct before powering on.
(2) Make sure to spend about 5-minute (or more) warming up time after powered on before starting main
operation.
10. Maintenance and adjustment
(1) Maintenance
This product rarely needs maintenance since it does not have any movable part in the detection part
due to its structure. However, depending on its usage status, appropriate maintenance is required.
Typically, regular check-ups twice a year is recommended.
The following is a check list for the regular check-ups for your reference.
[List of regular check-up items]
- Appearance
- Corrosion status of pressure inlet
- Insulation resistance between each terminal and the case (100 MΩ or more at 50 VDC)
- Re-tightening of connecting screw
- Output check by a reference pressure gauge
・ Avoid electrostatic-charge.
Cautions
・ DO NOT use thinner nor benzene which causes deterioration or failure.
sunlight, particles, salt or sulfur in the air.
When cleaning this instrument, use a soft damp cloth.