Prosonic M FMU 40/41/43 with HART and 4 ... 20 mA
8
Maintenance and repairs
8.1
Exterior cleaning
When cleaning the exterior of the Prosonic M, always use cleaning agents that do not
attack the surface of the housing and the seals.
8.2
Repairs
The Endress+Hauser repair concept assumes that the measuring devices have a
modular design and that customers are able to undertake repairs themselves.
Spare parts are contained in suitable kits. They contain the related replacement
instructions.
All the spare parts kits which you can order from Endress+Hauser for repairs are listed
with their order numbers in the section "Spare parts".
For more information on service and spare parts, contact the Service Department at
Endress+Hauser.
8.3
Repairs to Ex-approved devices
When carrying out repairs to Ex-approved devices, please note the following:
• Repairs to Ex-approved devices may only be carried out by trained personnel or by
Endress+Hauser Service.
• Comply with the prevailing standards, national Ex-area regulations, safety instructions
(XA) and certificates.
• Only use original spare parts from Endress+Hauser.
• When ordering a spare part, please note the device designation on the nameplate.
Only replace parts with identical parts.
• Carry out repairs according to the instructions. On completion of repairs, carry our the
specified routine test on the device.
• Only Endress+Hauser Service may convert a certified device into a different certified
variant.
• Document all repair work and conversions.
8.4
Replacement
After a complete instrument or electronic module has been replaced, the parameters
can be downloaded into the instrument again via the communication interface.
Prerequisite to this is that the data were uploaded to the PC beforehand using the
ToF Tool / Commuwin II. Measurement can continue without having to carry out a new
setup. Only a linearisation and a tank map (interference echo suppression) have to be
recorded again.
Maintenance and repairs
43