Advanced Menu; Decimal Point Shift; Ma Parameter; Filter Function - PCE Instruments PCE-SLT User Manual

Sound level meter
Table of Contents

Advertisement

10

Advanced menu

To access the extended menu, press the "SET" key for two seconds.
This menu is organised as follows:
dPSt

Decimal point shift

4-A4

mA parameter

20-A20
mA parameter
FiLt

Filter function

CtHY
Hysteresis for the control function
ALHY
Hysteresis for the alarm function
oFSt
Offset
GAin
Gain setting
Unit
Set RS232 unit
10.1
Decimal point shift
To move the decimal point, first press the "SET" key for two seconds. "dPSt" appears on the
display. Now press the arrow keys to access this configuration mode and to move the decimal
point. Press the "SET" key to save the setting.
10.2
4 mA Parameter
To change the parameterisation for 4 mA, first press the "SET" key for two seconds. "dPSt"
appears on the display. Press the "SET" key again. "4-A" now appears on the display. Now press
the arrow keys to access this configuration mode and to change the parameterisation for 4 mA.
Press the "SET" key to save the setting.
10.3
20 mA parameter
To change the parameterisation for 20 mA, first press the "SET" key for two seconds. "dPSt"
appears on the display. Now press the "SET" key twice. The display now shows "20-A". Now
press the arrow keys to access this configuration mode and to change the parameterisation for
20 mA. Press the "SET" key to save the setting.
10.4
Filter function
To change the parameterisation for the filter function, first press the "SET" key for two seconds.
"dPSt" appears on the display. Now press the "SET" key three times. "FiLt" now appears on the
display. Now press the arrow keys to access this configuration mode and to change the
parameterisation for the filter function. The higher the value, the more filtering takes place. Press
the "SET" key to save the setting.
10.5

Hysteresis for the control message

To change the parameterisation of the hysteresis for the control message, first press the "SET"
key for two seconds. "dPSt" appears on the display. Now press the "SET" key four times. "CtHY"
now appears on the display. Now press the arrow keys to access this configuration mode and to
change the parameterisation for the hysteresis. Press the "SET" key to save the setting.
© PCE Instruments
8