Page 1
INSTALLATION AND OPERATION MANUAL Traction table TRITON 6M Manufacturer MEDEN - INMED, Spółka z o.o. For service & repair: UL. WENEDÓW 2 Please contact your local DJO office or 75-847 KOSZALIN authorized dealer. POLSKA www.djoglobal.com TEL. +48 94 347 10 40 FAX.
2. FEATURES OF TRITON 6M TRACTION TABLE 2.1. USAGE 2.2. CE MARK 2.3. MARKING 2.4. TECHNICAL DATA 3. TRACTION TABLE TRITON 6M DESIGN AND FUNCTIONS 3.1. TABLE COMPONENTS 3.2. EQUIPMENT AND ADDITIONAL ACCESSORIES 3.3. TABLE SET 3.4. TRANSPORT AND STORAGE 4.
1. INTRODUCTION Congratulations on a perfect choice of our traction table. TRITON 6M is stable, reliable and easy to use. These features result from careful design, manufacturing methods applied in production and the quality assurance system. Reading and following instructions provided in this manual will ensure safe and long lasting use of our table.
2. FEATURES OF TRITON 6M TRACTION TABLE This manual describes the features of the TRITON 6M table, not all of the information and drawings contained in it may apply to your table. The manufacturer reserves the right to make changes, not depicted in the current version of the manual, which will not cause degradation of functional and safety conditions of the product.
Pelvic section angle adjustment - lowering Pelvic section angle adjustment - raising Lock/Unlock moveable bed section Head and Chest adjustment control lever Arm section adjustment lever Graduation range of the pelvic section 01-12-2017 TRITON 6M User Manual.doc Strona 5 z 27...
2.4. Technical data Table 1 shows the technical characteristics of the TRITON 6M traction table: TRITON 6M Technical data arms, chest and head section mid section pelvic section Table length leg section [cm] total 217 (leg section extended) Angle adjustement/...
The frame of the table is made of welded steel sections coated with powder paint. It includes the following components: Picture 1 - Elements of the TRITON 6M Traction table ①base frame the table support base, equipped with 4 wheels with total lock;...
CAUTION ! Use only accessories recommended or provided by the manufacturer. Every TRITON 6M traction table is provided with a breathing slot and reinforced mounting point for the traction belt in its front section. Additionally, the tables may be equipped with the following accessories: - traction unit TRU-TRAC/TRITON (recommended by Meden-Inmed);...
Do not use a table with a power cord other than provided by the manufacturer. CAUTION ! TRITON 6M traction table has a limited load capacity! The safe working load should not exceed 200 kg when lifting and the total load should not exceed 275 kg.
5. PREPARING THE UNIT FOR USE CAUTION ! When setting or changing the place of use of the TRITON 6M traction table, ensure that there are no objects in the space below the bottom frame of the table. Do not allow access to children or animals.
Select the place for the use of the table with a minimum of 80 cm clearance on each side of the unit. Picture 3 - Recommended positioning of the TRITON 6M Traction table Once the table is in position, unpack it from the transport foil and plug the power cord into the mains socket, which is freely accessible when using the table.
Before starting work the user should make sure that the patient is in a safe position. The height of the Triton 6M table is adjusted using the hand remote and foot control. To raise the table to the desired height, press on the foot control or the on the remote control.
Picture 6 - Regulation elements of TRITON 6M Traction table In order to adjust the tilt of the TRITON 6M front section, release the gas spring lock by holding down the lever (pic. 6 - position A), set the desired position of the front section, and lock in place releasing the lever.
(No 5). The traction unit base in the TRITON 6M Traction tables has a rotation of ± 90 degrees, as well as a step blockade every ~13 degrees. In order to unlock the unit base stand, unscrew the safety lock (pic.8 - A) - the safety knob is protected from being unscrewed fully, set the working degree of the...
Picture 8 – Attaching and replacing the traction unit base arm of the TRITON 6M Traction table. 6.5. Assembly of traction belt CAUTION ! During the use of the traction belt, refer to the User Manual enclosed by the belt manufacturer.
AC outlet and protected against accidental damage during relocation. CAUTION ! It is forbidden to move the TRITON 6M Traction table with a patient on the table. CAUTION ! Any manipulation of the position change lever WORK-TRANSPORT should be performed with the table without the patient.
Always shut down the table and prevent the table from being used by unauthorized persons. Then check it is locked, by testing the functions. The TRITON 6M is equipped with an access lock mechanism for the height adjustment function. There...
7.3. Checking the gas springs TRITON 6M Traction tables are equipped with gas springs for easy adjustment of the arms, chest & head and leg section inclination angle. To ensure proper performance of the gas springs check the springs visually for leaks every 3 months.
7.5. Periodic check of electrical safety The electrical safety check should be performed or ordered not less than once every 2 years, and after every fault or repair of the actuator and/or the power mechanism of the traction table. As a basic requirement, the review should include: - check for mechanical damage of wiring system of the table;...
8.0. TROUBLESHOOTING Failure indication Possible cause Action to be taken The height adjustment Exceeded the continuous work cycle If necessary, remove the patient from the mechanism does not work. table, with help if required. Remove the mains plug from the electrical socket and wait a minimum of 1 hour.
Essential Performance - there are no essential performance characteristics according to the risk assessment. *Traction table TRITON 6M Guidance and manufacturer’s declaration – electromagnetic emissions The equipment* is intended for use in the electromagnetic environment specified below. The customer or the user of the equipment* should assure that it is used in such an environment.
Page 23
IEC 61000-4-8 environment. NOTE U is the a.c. mains voltage prior to application of the test level. 01-12-2017 TRITON 6M User Manual.doc Strona 23 z 27...
Page 24
Otherwise, degradation of the performance of this equipment could result. NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies. 01-12-2017 TRITON 6M User Manual.doc Strona 24 z 27...
Page 25
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people. 01-12-2017 TRITON 6M User Manual.doc Strona 25 z 27...
17. This warranty, in the case of consumer sales, does not exclude, limit or suspend the buyer's rights resulting from the non-conformity of the goods with the sales contract. Traction table TRITON 6M Date, signature and warranty stamp: 01-12-2017 TRITON 6M User Manual.doc Strona 26 z 27...
Page 27
Repairs Registry User’s Notes 01-12-2017 TRITON 6M User Manual.doc Strona 27 z 27...