Deenergise device and secure against being switched on again

Extreme danger!
If work is to be performed on the devic-
es, they must be deener-gised. That
means that all live lines must be
switched off.
Switching off the device alone is insuf-
ficient, as certain points may remain
energised. Therefore, switch off fuses
before perform-ing any work and – if
possible – remove them.
Extreme danger!
Severe accidents can occur if the device is mistakenly switched
on again.
Immediately after deenergising, secure all switches or fuses
used for deenergising against being switched on again.
• Lock the fuse box with a padlock.
22
• In case of circuit breakers which cannot be
screwed out, a strip of adhesive tape may be ap-
plied over the actuation lever with the wording
"Do not switch, danger!"
01605 / 2
A prohibition sign with the text:
"Work in progress!"
• "Place: .............................."
• "Sign may only be removed by: ............"
must always be securely mounted immediately.
Extreme danger!
Prohibition signs must not be hung on or touch live parts.
01510 / 1
01511 / 1