Fault Finding - EMS 1000-3 Operation & Maintenance Manual

Advertisement

Electronic Micro Systems

Fault finding

In the unlikely occurrence of a fault, a circuit diagram is included.
The electronic components inside the system are protected by a thermal cut out switch set to
85°C, if the heater remains on for long periods of time, especially at high temperatures there
is a chance the internal temperature will reach this cut off value, it will automatically re-set
when cooled down.
The heater plate is a sealed unit and will have to be returned to EMS Base for service.
The temperature controller is an Ascon Tecnologic KR3 series, the full set up manual for the
controller is enclosed, further details can be found at
www.t-uk.co.uk
However please refrain from resetting the controller to "factory settings" as these are Ascon
settings and NOT ESM and it could be a long and laborious task to make operational again,
probably need returning to ESM !