TS 10 1 Introduction 1.1 Intended use This unit is intended for industrial use in gas analysis systems. It's an essential component for conditioning the sample gas to protect the analysis instrument from residual moisture in the sample gas. Please note the specifications in the data sheet on the specific intended use, existing material combinations, as well as pressure and temperature limits.
TS 10 2 Safety instructions 2.1 Important advice Operation of the device is only valid if: – the product is used under the conditions described in the installation- and operation instruction, the intended application according to the type plate and the intended use. In case of unauthorized modifications done by the user Bühler Technolo- gies GmbH can not be held responsible for any damage, –...
TS 10 2.2 General hazard warnings The equipment must be installed by a professional familiar with the safety requirements and risks. Be sure to observe the safety regulations and generally applicable rules of technology relevant for the installation site. Prevent malfunctions and avoid personal injuries and property damage.
TS 10 3 Transport and storage Only transport the product inside the original packaging or a suitable alternative. The equipment must be protected from moisture and heat when not in use. It must be stored in a covered, dry and dust-free room at a temperature of -20 °C to 60 °C (-4 °F to 140 °F).
TS 10 4 Installation and connection 4.1 Installation site requirements The unit is only intended for wall-mounted use in enclosed areas. Adequate protection from the weather must be provided when used outdoors. Install the unit leaving enough room below the cooler to discharge the condensate. Leave room above for the gas supply. Be sure to maintain the approved ambient temperature.
Page 9
TS 10 CAUTION Wrong mains voltage Wrong mains voltage may damage the device. Regard the correct mains voltage as given on the type plate. Plug connection This device has one DIN 43650 plug each for the power supply. If the lead is connected correctly, these cannot be confused. Therefore please be sure to correctly reassemble the plugs after connecting the wires.
TS 10 5 Operation and control NOTICE The device must not be operated beyond its specifications. After switching on the cooler, both the pump and fan will start. Please refer to the data sheet or technical data for performance data and maximum ratings. Bühler Technologies GmbH BE450009 ◦...
TS 10 6 Maintenance The basic version of the cooler requires no special maintenance. However, it may have different options depending on the cooler model. In this case the following maintenance must be per- formed regularly: – Optional peristaltic pump: Check hoses During maintenance, remember: –...
TS 10 7 Service and repair This chapter contains information on troubleshooting and correction should an error occur during operation. Repairs to the unit must be performed by Bühler authorised personnel. Please contact our Service Department with any questions: Tel.: +49-(0)2102-498955 or your agent If the equipment is not functioning properly after correcting any malfunctions and switching on the power, it must be inspected by the manufacturer.
TS 10 7.3 Replacing the hoses of the peristaltic pump (option) – Turn off gas supply. – Switch the device off and disconnect power supply. – Remove the supplying and draining hoses from the pump (Take care of the safety instructions!). –...
TS 10 8 Disposal The applicable national laws must be observed when disposing of the products. Disposal must not result in a danger to health and environment. The crossed out wheelie bin symbol on Bühler Technologies GmbH electrical and electronic products indicates special disposal notices within the European Union (EU).
TS 10 9 Appendices 9.1 Technical Data TS 10 precooler Technical Data Ambient temperature: 0 .. 60 °C Electric supply: 115 V, 60 Hz or 230 V, 50 Hz Power input: 25 W Type of protection electric: IP 20 Housing: Stainless steel Installation: wall-mounted...
Page 19
RMA-Formular und Erklärung über Dekontaminierung RMA-Form and explanation for decontamination RMA-Nr./ RMA-No. Die RMA-Nr. bekommen Sie von Ihrem Ansprechpartner im Vertrieb oder Service. Bei Rücksendung eines Altgeräts zur Entsorgung tragen Sie bitte in das Feld der RMA-Nr. "WEEE" ein./ You may obtain the RMA number from your sales or ser- vice representative.