Section V
Model 400E/EL
Paragraph 5-19 to 5-21
5-19. INPUT IMPEDANCE
CHECK.
5-20. Input Resistance Check.
a.
Connect the 50 Q output of the test oscillator to the input of the
400E/EL.
b.
Set the test oscillator and the 400E/EL to the 3 volt range.
Set
the oscillator output to 40 cps (Hz) and adjust the output for a full
scale indication as shown.
c.
Connect a 100 K Q resistor between the test oscillator output and
the 400E/EL input as shown in Figure 5-2.
hp 400E/EL
TEST OSCILLATOR
AC VOLTMETER
hp 651A
1OOKN £19
O-<
&
On
400 /EL-8-0474
thee — 6|
hpioii0a
Figure 5-2.
Input Impedance Check
d.
The 400E/EL
indication should not drop more than one small
scale division from full scale.
This verifies an input resistance
of 10 M 2.
5-21. Input Capacity Check (Option 01, 02, and Standard 400E/EL).
a.
Connect a test oscillator,
a 100 K © resistor,
and the 400E/EL
as shown in Figure 5-2.
b.
With the 400E/EL on the 3 volt range,
adjust the test oscillator
for a full scale reading on the 400E/EL at 400 cps (Hz).
c.
Increase the test oscillator frequency until the 400E/EL indication
drops to 2.12 volts.
This should occur at a frequency of 105 Kc
(KHz) or greater, verifying an input capacity of 9 pf or less on the
3 volt range.
d.
Repeat steps a and b with the 400E/EL on the 1 volt range.
e.
Increase the test oscillator frequency until the 400E/EL indication
drops to 0.707 volts.
This should occur at a frequency of 55 Kc
(KHz) or greater, verifying an input capacity of 24 pf or less on
the 1 volt range.
5-8