HP 400E Operating And Service Manual page 26

Ac voltmeter
Hide thumbs Also See for 400E:
Table of Contents

Advertisement

Section V
Model 400E/EL
Paragraphs 5-12 to 5-13
RANGE
DIGITAL VOLTMETER
SELECTOR
hp 3439A
hp 3443A
DC STANDARD/ DIFFERENTIAL
VOLTMETER
hp 740A
DC NULL
hp 400E/EL
VOLTMETER
AC VOLTMETER
hp 419A
TO
O-IOMV
OC OUTPUT
one
SUPPLY
ON REAR
Fi
|
|
|
|
|
THERMAL CONVERTER
\
I
|
Ap 11049A (3v)
/Z bpiiosoa (iv)
OUTPUT CABLE | f
pPilOS5A
|
TEST OSCILLATOR
hp 651A
400E/EL-C-0473
Figure 5-1.
Accuracy and Frequency Response Test Setup
5-12. Reference Supply Calibration.
5-13. Use the following procedure to calibrate the thermal converter and reference
supply.
a.
Connect the dc standard (740A), the 400E/EL,
the null voltmeter (419A),
the reference supply, and a 3 volt thermal converter (11049A) as shown
in Figure 5-1.
NOTE
If a 400E/EL Option 02 instrument is used,
set the
REL.
REF adjustment to the fully clockwise ABSO-
LUTE position before making accuracy check.
The reference supply and the thermal converter are sensitive to varia-
tions in ambient temperature,
Ensure that the ambient temperature
variations are less than +2.0°C.
b.
Set the de standard output to +3.000 voits dc.