AFFINITY (WIRELESS PROBE)

To prevent fire and explosion:
Do not open in a hazardous or potentially explosive area.
Eliminate all ignition sources; such as pilot lights or cigarettes.
Only use AA Type Energizer
Do not attempt to charge non-rechargeable batteries, for example, the Affinity batteries.
iQV11
Various peripherals can be connected to Affinity to measure different parameters, but these peripherals may not
be certified or certified appropriately for use in potentially explosive atmospheres. It is the responsibility of the
operator to determine the protection concept and classification of equipment required for a particular
application and whether these peripherals are suitable for use in a potentially explosive atmosphere.
The instrument has been designed to be used with peripheral accessories that can be connected to the three
connection sockets on Affinity, namely Temperature Probe, Anemometer, and Humidity Probe. The entity
parameters for these sockets is as follows:
Connector
A – Temperature Probe
B – Humidity Probe
C – Anemometer Probe

CONNECTION

Affinity connects to the i-Q Series by Bluetooth. To turn Affinity on, press and hold the power key, see
Annotation 2 on
Figure 5 –
OMiQGEM-L Issue.02
Page 16 of 51
®
L91 batteries.
Uo
Io
Po
5V
5mA
7mW
8.6V
16mA
33mW
8.6V
3mA
6mW
Affinity, for two seconds. Pairing is done through the i-Q instrument.
Ci
Li
Co
0µF
0mH
500µF
2µF
0mH
499µF
0µF
0mH
500µF
Lo
500mH
500mH