placed between the jaws of the electrode adapter and
remain flat. Do not attempt to place the jaws of.the
electrode adapter so close to the circular part of the
electrode that the tab of the electrode is bent or contact
is made with the conductive gel. Gently tug on the
electrode adapter to ensure that the electrode adapter is
properly placed on the electrode.
Good and accurate placement on the first attempt
should be your goal for each electrode. Each time an
electrode is lifted off the skin and attached again, the
conductive gel becomes weaker and less effective.
Understa
------
--
Jnding
When a Signal is
Acquired
Your PageWriter 100 cardiograph attempts to acquire
a.good signal for an Auto report before you press the
(Auto) key. Hewlett-Packard cab this Pre-acquisition.
Pre-acquisition is activated when the cardiograph is
turned on and remains active until an Auto report
begins to print. Pre-acquisition is deactivated by
printing an Auto report to allow for copies of the Auto
report to be printed (see "Making Copies of Auto
ECGs"). Pre-acquisition is also deactivated whenever an
electrode is disconnected.
Pre-acquisition is reactivated when a Manual report is
finished printing.
When Pre-acquisition is active, it is important for the
patient to stay still and relaxed. This will help ensure a
good signal is captured prior to printing an Auto report.
Note
I
VI
Pre-acquisition is not used for Manual ECG reports.
Manual ECG reports display ECG data in real-time.
Recording an ECG 2-3