Download Print this page

Scale-Tronix 5002 Operating And Service Manual page 17

Second generation weighing systems

Advertisement

pushbutton again. The software version will be displayed
(ei. "YU 1.42.2).
This is provided as a reference aid in
determining which features
are contained within the
software.
2)
Press the "ST'
pushbutton.
The
scale will
display "dAtE" (date). Again press the "ST" pushbutton.
The release date of the software will be displayed.
A
display of "03.26.96" would correspond to March 26th,
1996. This is provided as a reference aid in determining
which features are contained within the software.
3)
Press
the
"ST"
pushbutton.
The
scale
will
display "SCALE".
Press the "ST" pushbutton again; the
model
number
of the
scale
will be
displayed
(i.e:
"5002", "6002", etc.).
This is provided as a reference
aid to help with customer trouble-shooting and is set at
the factory.
4)
Press the "ST" pushbutton.
If the scale's serial
number has been programmed at the factory the scale
will display "SEr no." (serial number).
Press the
"ST"
pushbutton again; the serial number of the scale will be
displayed (e.i. "10563"). This is provided as a reference
aid to help with customer trouble-shooting and is set at
the factory.
Newer versions of the software (10/07/95
or later) skip these displays if the serial number has not
been entered at the factory.
5)
Press the
"ST" pushbutton.
The scale will
display "AutOFF" (automatic shut-off time). Again press
the "ST" pushbutton.
The currently selected number of
seconds before the scale shuts-off will be displayed.
This is normally set at the factory to be 45 seconds (120
seconds for the model 6102).
lt can be altered in 15
second increments fram 15 seconds to 180 seconds (3
minutes).
An additional "Conf (continuous) option is
provided for some special applications; this prevents the
scale from automatically shutting-off; the scale can then
only be turned off by pressing and holding the "ON"
pushbutton.
Use of the continuous mode of operation
is not recommended since it will discharge the battery.
Portable scales equipped with the optional power supply
(see
section
2.90)
may
have
this
option
set
to
continuous
since
there
is no
problem
with
battery
discharge.
Ta alter the length of the automatic shut-off time press
the "RE CALL" pushbutton. The automatic shut-off time
will advance through the available options.
Stop at the
desired time interval.
Line
powered
scales
which
remain
on
continuously,
such as the model §102 physician's office scale and the
mode!
6202
ramp
scale, should
display
"AC
ОРГ,
indicating "AC operation".
This is programmed
at the
factory and cannot be changed in the "set-up" mode.
6}
Press the "ST" pushbutton.
Newer versions of
the scale will display "AC con" (AC power continuous
operation),
Older
versions
of the
scale
(software
versions below 2.00) will not provide this option, skip to
paragraph 7 below.
This feature is used in the model
6102 built-in scale to cause the scale to remain "on" as
long as AC power is provided
(automatic shut-off will
not occur).
Pressing the "ST" pushbutton again will show this option
selected as "On" or "OFF".
To alter the selection press
the "RE CALL" pushbutton.
This option should remain
set at "OFF" except for the model 6102 built-in scale.
7)
Press the "ST"
pushbutton.
The
scale will
display
"YES"
(resolution
of weight),
Press the "ST"
pushbutton again. The "POUNDS" and "KILOGRAMS"
annunciators
will
light
and
two
numbers
will
be
displayed.
These
numbers
represent
the
selected
resolution
of the scale.
Pressing
the
"RE
CALL"
pushbutton will allow the following selections:
0.1 pounds
/
0.05 kilograms
0.1 pounds
/
0.1 kilograms
0.2 pounds
/
0.1 kilograms
0.5 pounds
/
0.2 kilograms
1,0 pounds
/
0.5 kilograms
Normal factory selection is 0.1 pounds / 0.1 kilograms.
Since
the
scale
uses
the
selected
resolution
to
determine weight lock-in, use of a larger resolution will
reduce waiting time, particularly for unsteady patients.
8)
Press
the
"ST
pushbutton,
The
scale
will
display
"UnitS"
(weight
units)
Press
the
"ST"
pushbutton again. The "POUNDS" and "KILOGRAMS"
annunciators will light according to the selected weight
units.
Option
choices
of
"POUNDS"
and
"KILOGRAMS", "KILOGRAMS" only and "POUNDS"
only
may
be
selected
by
use
of the
"RE
CALL"
pushbutton.
If a particular weight unit is disabled, its
front panel pushbutton will no longer operate and may
be covered with a strip of tape to prevent confusion.
9)
Press the "ST"
pushbutton.
The
scale will
14
$002, 5102, 5602, 6002, 6102, 6202, 6502, 6702

Advertisement

loading

Related Products for Scale-Tronix 5002

This manual is also suitable for:

560251026002670261026202 ... Show all