Temperature Control; Flanking I.v. Pole Brackets; Unpacking And Set Up - level 1 System 1000 Operator's And Service Manual

Fluid warmer 110v
Table of Contents

Advertisement

connections and can be easily changed out without patient disconnection, or
repriming of the entire disposable set. The Filter with Gas Vent should be
changed every three hours or when the filter becomes clogged, which ever
comes first.
|
|
|
Temperature Control
The System 1000 employs a safe circulating water heating system, inherently
free of any "hot spots." The primary temperature control circuit sets the
circulating water bath to a temperature of approximately 42°C for efficient heat
exchange and maximum fluid warming.
Flanking I.V. Pole Brackets
The sturdy design of the System 1000 offers a compact, custom I.V. pole
mount set up. This versatile workstation features flanking side poles for
SIMS Level 1 pressure infusers, syringe pumps or infusion pumps.
Unpacking
and Set Up
The System 1000 packaging should contain the following items:
Contents
Quantity
H-1000 Warming Unit
1
H-1000 1.V. Pole Base
1
H-1000 1.V. Bag Hanger
1
H-1000 1.V. Pole with
Flanking Brackets
Accessory Pack:
H-1000 Manual
D-series Instruction Card
Y-Connector
Black Pneumatic Tubing
Hex Wrench
1. Place the base upright on its wheels.
Lock casters to prevent pole movement
during set up.
2. Remove the three threaded pole screws on the vertical base tube.
Slide the
UV. pole with flanking brackets down over the vertical base tube lining up the
three screw holes in the pole with the three screw holes in the base as shown
in Figure 2. Noite: the flanking brackets shouid face the longer side of the
base.
3. Thread the three pole screws back through the holes at the base of the pole
and tighten.
.
|
|
4. Attach the System 1000 warming unit to the flanking brackets by sliding the
— eight threaded screws on the back of the unit into the eight keyhole notches
on the flanking I.V. pole brackets.
(Refer to Figure 2.) Tighten all eight
screws to secure warmer to |.V. pole and flanking brackets.
o. Lock
I.V. bag hanger into place on the top of the I.V. pole.