Fluid warmer 110v
Table of Contents

Advertisement

2. CLEAN FAN FILTER
Remove the black plastic fan filter from the bottom of the unit. It is
located "inside of " the four fan guard screws and snaps in and out.
Clean it with warm soapy water, allow it to dry, and snap it back into
place.
Maintenance and Testing Log
All maintenance and testing should be done by qualified personnel.
Every 30 days:
Change Distilled water
Lubricate O-Rings
Test Alarm Switch
Every 12 months:
Replace O-Rings
Clean Fan Filter
Test Add Water Alarm
Test Disposable Alarm
Test Over Temp Alarm".
Performance Test/ Verify
Temperature Calibration
Service
NOTICE:
ALL SERVICE MUST BE PERFORMED BY SIMS LEVEL 1, INC. OR
ITS AUTHORIZED AGENTS.
SERVICE BY OTHERS VOIDS THE WARRANTY
AND TRANSFERS LIABILITY FOR MALFUNCTIONS OF THE DEVICE TO THE
SERVICING ORGANIZATION.
Warranty Service
Units received for repair which have not been obviously abused or impact
damaged and still under warranty will be promptly repaired and returned at no
charge. Contact your authorized distributor, or the SIMS Level 1 Service
Department before shipping the unit to SIMS Level 1 for repair. The Service
Department will issue a Return Goods Authorization Number (RGA #). The
Service Department will need the SERIAL NUMBER of the unit and a description
of the problem.
This SIMS Level 1 Fluid Warmer is warranted by SIMS Level 1, Inc. to be free of
material and workmanship defects for a period of 1 year (12 months) from the
date of purchase, abuse and impact damage excluded. SIMS Level 1 reserves
the right to replace any or all components in lieu of repair. During the period this
22