Mechanical Measurements; Automatic Recording Of Accelerometer Frequency Response - BRUEL & KJAER 1013 Instructions And Applications

Beat frequency oscillator
Table of Contents

Advertisement

phones) should be small in comparison with the time constant which deter-
mines the compressor speed.
Fig. 5.17 shows a frequency characteristic of a piezo-electric hydrophone
measured by the previously mentioned system.
Exciting Section
r----------------------------------------------,
@o
~-~~
i
@) • . . .
~-+
:
-----,
~~
··
1013
AO 0014
Accelerometer
under test
Cathode
Follower
t
'
I
l
'
I
'
I
I
I
I
I
I
I
I
I
I
2804
I
.
I
Control voltage to regul. am pl.
1
I
- - - - - - - - - - - - - - - - - - - - - - - - -
________________________ J
Mechanical Connection
2804
Fig. 5.18. Measuring arrangement for frequency response calibration of
accelerometers
.
MECHANICAL MEASUREMENTS
Automatic
Recording of
Accelerometer Frequency
Response.
In the production of high quality electromechanical transducers it is a
important asset to deliver wit hevery unit an individually recorded amplitude
versus frequency characteristic.
The B
&
K accelerometers, which are duly supplied with individually re-
corded amplitude vs. frequency curves, are measured on a measuring
arrangement equivalent to the one shown in Fig. 5.18. The arrangement
con-
40