Troubleshooting; Contact Information - WatchGas SST Dock Quick Start Manual

Hide thumbs Also See for SST Dock:
Table of Contents

Advertisement

TROUBLESHOOTING

Detector fails bump, fresh air, or calibration:
1.
If fresh air fails, verify the filter is not clogged and the air at the inlet is not contaminated.
2.
Check that gas bottles are connected, sufficiently pressurized, and contain the right gas
concentration. (information on the bottle)
3.
Verify the station function with a known functional detector.
Detector fails alarm test:
1.
Redo the test
2.
Manually inspect the detector for LED, vibration, and audio alarms
3.
Test a known fully functional detector in the station to verify station function
Station fail:
1.
Clear fault and retry
2.
Contact service representative

CONTACT INFORMATION

Klaverbaan 121, 2908 KD, Capelle a/d IJssel, the Netherlands
info@watchgas.com | www.watchgas.com
313 N. State Hwy 342, Red Oak, Texas 75154, United States
info@watchgasusa.com | www.watchgasusa.com
SST Dock | QSG | V 1.0 | EN | 15-06-23
WatchGas is dedicated to continuously improving its products. Therefore, the specifications and features mentioned in this QSG are subject to change without prior notice.
© 2023 WatchGas B.V.
11

Advertisement

Table of Contents
loading

Table of Contents