Calibration

The bq27421EVM must be calibrated to ensure accurate value reporting. This is done using the
Calibration window in bqStudio
8.1

Calibrating the bq27421

1. Select the types of calibration to be performed (see
2. Enter the measured values for the types selected.
3. Press the button to calibrate.
8.2

Voltage Calibration

Voltage calibration usually is not required. If needed, follow these steps:
1. Measure the voltage across Pack+ and Pack–.
2. Type the voltage value in mV into Enter measured value.
3. Press the Calibrate button.
8.3

Board Offset Calibration

This performs the offset calibration for the current offset of the board.
It is expected that no current is flowing through the sense resistor while performing this calibration step.
1. Remove load and short PACK– to LOAD–.
2. Press the Calibrate button.
SLUUA63A – April 2013 – Revised July 2016
Submit Documentation Feedback
(Figure
11).
Figure 11. Calibration Screen