Download Print this page

KERN 20K10 Operating Instructions page 2

Hanging scale

Advertisement

4
Transport and storage
4.1
Testing upon acceptance
When receiving the appliance, please check packaging imme-
diately, and the appliance itself when unpacking for possible
visible damage.
4.2
Packaging
Keep all parts of the original packaging in case you need to
return the appliance.
Only use original packaging for returning.
5
Unpacking, Setup and Commissioning
5.1
Installation Site, Location of Use
The balances are designed in a way that reliable weighing
results are achieved in common conditions of use.
You will work accurately and fast, if you select the right location
for your balance.
Therefore, observe the following for the installation site:
Avoid extreme heat as well as temperature fluctuation
caused by using next to a radiator or in the direct sunlight;
Protect the balance against direct draughts due to open
windows and doors;
Avoid jarring during weighing;
Protect the balance against high humidity, vapours and
dust;
Do not expose the device to extreme dampness for longer
periods of time. Non-permitted condensation (condensa-
tion of air humidity on the appliance) may occur if a cold
appliance is taken to a considerably warmer environment.
In this case, acclimatize the appliance for ca. 2 hours at
room temperature.
Avoid static charging of the material to be weighed, weigh-
ing container and windshield.
If electro-magnetic fields or static charge occur, or if the power
supply is unstable major deviations on the display (incorrect
weighing results) are possible. In that case, the location must
be changed.
5.2
Unpacking
Take the balance carefully out of the package and remove
the plastic cover.
5.3
Scope of delivery
Suspended balance
Batteries
Operating Manual
5.4
Battery operation
Push out the battery cover on the balance backside. Insert 3
AA batteries. Reinsert the battery cover. If the batteries are
empty, in the balance display appears "LO". Press ON/OFF-
key and replace the batteries immediately. In order to save the
battery, the balance switches automatically off after 4 minutes
without weighing. Deactivate this Auto-OFF function in the
menu as follows:
5.5
The menu
When the balance is switched off, press the HOLD-key
and keep it pressed
Do not release the HOLD-key. Press also the
ON/OFF/TARE-key and keep it pressed
Keep the ON/OFF/TARE-key pressed, however release
the HOLD-key
Press HOLD-key anew
Keep both keys pressed until in the display "tr" appears
HCB-BA-e-0732
Release both keys. The balance is situated in the menu.
Using the ON/OFF/TARE button you can select between
the following functions:
tr (zero tracking) : ON/OFF
AF (Auto off): ON/OFF
rST (Reset to factory setting): YES/NO
Confirm with HOLD-key
In the display appears the current setting „ON" or „OFF"
resp. „YES" or „NO".With help of the ON/OFF/TARE-key
you can select between „ON" or „OFF" resp. „YES" or
„NO". Confirm your selection with the HOLD-key. After a
short period the balance returns automatically to weighing
mode.
6
Adjustment
As the acceleration value due to gravity is not the same at
every location on earth, each balance must be coordinated - in
compliance with the underlying physical weighing principle - to
the existing acceleration due to gravity at its place of location
(only if the balance has not already been adjusted to the loca-
tion in the factory). This adjustment process must be carried
out during the initial start-up, after change in location and varia-
tion of surrounding temperature. To receive accurate measur-
ing values it is also recommended to adjust the balance peri-
odically in weighing operation.
6.1
Adjustment
With an adjustment weight, the weighing accuracy can be
checked and re-adjusted at any time.
Procedure when adjusting:
Observe stable environmental conditions. A short warming up
time of ca. 1 minute is recommended for stabilization. Switch
on balance Press the Unit-key and keep it pressed (approx. 3
sec), in the display first appears shortly „CAL ", after that there
appears the exact size of the adjustment weight. Suspend the
adjustment weight, a short time later appears „ F", the balance
returns automatically to the weighing mode.
In case of an adjustment error or incorrect adjusting weight the
display will show "E"; repeat adjustment process.
KERN
20K10
Adjustment
10 kg
weight
(M3)
(not added)
KERN
99K50
Adjustment
50 kg
weight
(M3)
(not added)
7
Operation
7.1
Weighing
• The balance is switched on by pressing the „ON/OFF/TARE"
key.
• The balance displays „0.0".
• Suspend material to be weighed.
• The weighing value appears in the display:
• If the material to be weighed is heavier than the weighing
range, the display will show "E" (=Overload).
• The balance is switched off by pressing the „ON/OFF" key a
longer time.
20K50
50K20
50K100
10 kg
20 kg
20 kg
(M3)
(M3)
(M3)
100K200 200K100 200K500
50 kg
100 kg
100 kg
(M3)
(M3)
(M3)
7

Advertisement