Operation
OUT/
OUT2
Min. distance between SETL and SETU: 20 °C/K (36 °F)
16
Output/output 2 (optional)
Value for 4 mA
(LRV)
Value for 20 mA
(URV)
Temperature
applied for 4 mA
(LRV)
Temperature
applied for 20 mA
(URV)
Error current
TTR31, TTR35
-50 to 130 °C
(-58 to 266 °F)
Lower range value in
increments of 0.1 °C
(0.18 °F)
-30 to 150 °C
(-22 to 302 °F)
Enter upper range value in
increments of 0.1 °C
(0.18 °F)
Take temperature value as
lower range value (not via
PC software)
Take temperature value as
upper range value (not vial
PC software)
Current value in event of
error:
MIN = ≤ 3.6 mA
MAX = ≥ 21.0 mA
HOLD = last value