Swan AMI Oxytrace
Quick Reference Guide
Maintenance
6 monthly
Yearly
Every 2
nd
year
How to remove Oxytrace Sensor
Reorder
codes
Error
Messages
Swan Analytical
Australia Pacific
www.swan-analytical.com.au
• Clean the Oxytrace G membrane with a soft tissue.
• Clean the Faraday electrode (if installed)
• If necessary, replace filling electrolyte. If the sensor is exposed to air frequently and
during long time intervals, the electrolyte and membrane may have to be changed
earlier.
• Clean flow cell and flow meter, if dirty
• Replace Swansensor Oxytrace G membrane by using a new, pre-filled sensor cap.
Oxytrace
Maintenance
Video
A – Fixing sleeve
B – Sensor
C – Flow cell
Unscrew the fixing sleeve [A].
Remove the sensor from the flow cell.
Note: A vacuum may form when the sensor is pulled out. Do not use force!
The sensor can be easily removed if the flow control valve is slightly opened.
Product Code
Description
A-87.290.050
Sensor Cap Oxytrace G
To view active error messages, press ENTER to open menu system, select
MESSAGES -> PENDING ERRORS. Consult the manual for a full list of error
messages.
HINT: With any Swan system, press ENTER 3 times from the home screen to bring
up the PENDING ERRORS list.