Pressure Module - Parr Instrument 4838 Operating Instructions Manual

Reactor controller
Table of Contents

Advertisement

Pressure Module

Press return key and release
Operation
Select
Mode
type/value
r-S
Run
SP
To be deter-
mined**
AL1H
0
LoC
OFF
** If units are psi, SP = 0
If units are Bar, SP = 1
If units are MPa, SP = 2
Press and hold down "SET" for 5-sec
Operation
Select
Mode
type/value
InPt
V5
tP-H
Determined
by transduc-
er**
tP-L
0
CTRL
ON/OFF
S-HC
Heat
ALA1
2
SALA
OFF
CoSH
ON
C-Sl
RTU
C-no
3
bPS
9600
Len
8
PrtY
Even
StoP
1
Comment
Run/Stop
Decimal point position
Upper-limit alarm 1
Lock mode (lock all keys or
only up/down arrow able
to use)
Comment
Input type (V5 = 0-5V, V10
= 0-10V, nA0 = 0-20mA,
nA4 = 4-20mA, nV =0-
50mV)
Upper-limit pressure range
Lower-limit pressure range
Control mode (ON/OFF ,
MANUAL, PID and PID-
PROG)
Heat/Cool control
Alarm operates when PV
value is higher than SV
Value + ALH setting
System alarm
Communication write
function that able to use
set point from software
Format type
Communication address
Communication baud rate
Data length setting
Parity bit setting
Stop bit setting
** Set according to the following:
Transducer
tP-H for
range - psi
psi
200
200
300
300
500
500
1000
1000
2000
2000
3000
3000
5000
5000
7500
7500
Press "SET" and release
Operation
Select
Mode
type/value
HtS
0
tPoF
0
4838 Reactor Controller
tP-H for
tP-H for
Bar
MPa
13.79
1.38
20.68
2.07
34.47
3.45
68.95
6.89
137.9
13.79
206.8
20.68
344.7
34.47
517.1
51.71
Comment
Heating hysteresis setting
Temperature correction
19