Features
This is a Digital Multi Meter providing most portability by means of stowing the instrument body together with probe in the notebook-size cover.
- Designed to international safety standards.
IEC 61010-1 Measurement CAT.Ⅲ 300V
Pollution degree 2
IEC 61010-031 (probe assemblies) - REL function to check the difference of measured values
- Auto power off function to save battery consumption
- Data hold function
- Diode and Continuity check function
- Auto-ranging function
- Frequency measurement function
- DUTY measurement function
(Express Pulse width / Pulse period as a percentage)
Specification
- Measuring ranges and accuracy
(23±5℃, under the 45%~75%RH)
DCV Function(5 Autoranging):Input impedance approx. 10MΩ
Range | Measuring range | Accuracy |
400mV | 0~600V | ±0.8%rdg±5dgt |
4V |
40V |
400V |
600V | ±1.0%rdg±5dgt |
ACV Function(4 Autoranging): Input impedance approx. 10MΩ
Range | Measuring range | Accuracy |
4V | 0~600V | ±1.3%rdg±5dgt (50/60Hz) |
40V | ±1.7%rdg±5dgt (~400Hz) |
400V | ±1.6%rdg±5dgt (50/60Hz) |
600V | ±2.0%rdg±5dgt (~400Hz) |
Resistance Function (6 Autoranging)
Range | Measuring range | Accuracy |
400![]() | 0~40M![]() | ±1.0%rdg±5dgt |
4k![]() |
40k![]() |
400k![]() |
4M![]() |
40M![]() | ±2.5%rdg±5dgt |
Diode check/Continuity check Function
Function | Measuring range |
Diode check | Test current approx. 0.4mA |
Continuity check | Buzzer beeps below about 120![]() |
Capacity Fuction (6 Autoranging)
Range | Measuring range | Accuracy |
4nF | ~200uF | ±5.5%rdg±10dgt |
40nF | ±3.5%rdg±10dgt |
400nF | ±3.5%rdg±5dgt |
4uF |
40uF |
200uF | ±4.5%rdg±10dgt |
Frequency (6 Autoranging)/ DUTY Function
Range | Measuring range | Accuracy |
10Hz | ~200kHz Input sensitivity: more than 1.5V (RMS) | ±0.1%rdg±5dgt |
100Hz |
1000Hz |
10kHz |
100kHz |
200kHz |
DUTY | 0.1~99.9% (Pulse width/Pulse period) | ±2.5%rdg±5dgt |
Note:
*At Voltage function, the Auto-ranging function is released by pressing the SELECT key. To measure a voltage again, turn the Function switch to the "OFF" position once. Then set it to the Voltage function again.
- Standards:
IEC61010-1 Measurement CATⅢ 300V, pollution degree 2
Measurement CATⅡ 600V, pollution degree 2
IEC61010-031
IEC61326 - Method of operation: △∑method
- Indication:
LCD maximum value 3999 (ACV, DCV, Ω, F) units, symbols - Over range display:
"OL" symbol is displayed on the LCD.
In case that the value is beyond effective measuring range at the position of Ω function range and manual range. - Auto-ranging:
Range shifts to upper range when indicated value is more than 3999. Range shifts to lower range when indicated value is less than 360. - Sampling rate: approx. 400ms
- Operating Environmental conditions
- indoor use
- altitude up to 2000m
- Temperature & Humidity range(guaranteed accuracy):
23℃±5℃ Relative humidity: less than 75% - Operating Temperature & Humidity range:
0℃~+40℃ Relative humidity: less than 80% - Storage Temperature & Humidity range:
-20℃~+60℃ Relative humidity: less than 70% - Insulation Resistance:
It should be more than 10MΩ/DC1000V between electrical circuit and enclosures. - Withstand Voltage:
It should be more than AC3700V/ for one minute between electrical circuit and enclosures. - Overload Protection:
Voltage function: 720V(RMS.) 10seconds
Resistance function: 250V(RMS.) 10seconds
Diode / Continuity: 250V(RMS.) 10seconds
Capacity function: 250V(RMS.) 10seconds
Frequency function: 250V(RMS.) 10seconds - Dimensions: approx. 107(L) x 54(W) x 10(D) mm
- Weight: approx. 70g(including batteries)
- Power source: Two LR44(SR44)1.5V or equivalent
- Accessories: Two LR44 (SR44) batteries
Portable holder
Instruction manual
![]()
The voltage shown above is the overload protection (overvoltage protection) for the instrument.
Make sure not to exceed the value of voltage shown above.
Instrument Layout
![KYORITSU - KEW 1018 - Instrument Layout Instrument Layout]()
- Function Selector Switch
- Display
- SELECT Key
- DATA HOLD Key
- Test Leads
Preparation
5-1 Checking Battery Voltage
Set the Function Selector Switch to other positions except the OFF position.
Battery Voltage is enough if indication is clear and BT symbol is not indicated in this bout.
If BT symbol is indicated or no indication on the display, follow to the Battery Replacement procedures shown in item9 in this document and replace with new batteries.
Measurements
Voltage Measurement (DCV, ACV)
![]()
- To avoid the danger of getting electrical shock, never make measurement on a circuit over 600V AC/DC. (electrical potential to ground 300V AC/DC)
- Do not operate Function Selector Switch during measurement.
- Do not make measurement when opening the instrument case.
DC Voltage Measurement (DCV)
- Set the Function Selector Switch to "
" position. (Then, "AUTO","
"and"mV" symbols are indicated on the display.) - Connect the black test lead to the negative side ofthe circuit under the test and the red test lead to the positive side of the circuit, then the measured value is indicated on the display. If you connect the test leads the other way, "-" symbol is indicated on the display.
AC Voltage Measurement (ACV)
- Set the Function Selector Switch to "
" position. (Then, "AUTO", "
", and "V" symbols are indicated on the display.) - Connect the test leads to the circuit under test. Measured value is indicated on the display.
Note) Even if short-circuit the input line at the range of AC4V, 2~5dgt may remain indicated.
Resistance Measurement
![]()
- To avoid the danger of getting electrical shock, never make measurement of the circuit in which electric potential exists.
- Do not make measurement when opening the instrument case.
- Set the Function Selector Switch to "Ω" position.(Then, "AUTO" and "MΩ" symbols are indicated on the display.) Make sure that the "OL" symbol is indicated on the display at this bout, then short the test lead tips and check "0" is indicated on the display.
- Connect the test leads to both ends of the resistanceunder test.
Measured value is indicated on the display.
Note) Even if short the test lead tips, indicated value may not be "0". But this is because of the resistance of test leads and not a failure.
Continuity Check/ Diode Check
![]()
- To avoid the danger of getting electrical shock, never make measurement of the circuit in which electric potential exists.
- Do not make measurement when opening the instrument case.
Continuity Check
- Set the Function Selector Switch to "
/
" position.
(Then, "
" and Ω symbols are indicated on the display.) - Make sure that the "OL" symbol is indicated on thedisplay at this bout, then short the test lead tips and check "0" is indicated on the display and check if the buzzer beeps.
- Connect the test leads to both ends of the resistanceunder test.
Measured value is indicated on the display. The buzzer beeps below about 120Ω.
Note) Even if short the test lead tips, indicated value may not be "0". But this is because of the resistance of test leads and not a failure.
Diode Check
- Set the Function Selector Switch to "
/
" position. (Then, "
" and "Ω" symbols are indicated on the display.) - Press the SELECT Key twice and set the instrumentto Diode Check mode. (Then, "
" and "V" symbols are indicated on the display.) Make sure that the "OL" symbol is indicated on the display at this bout, then short the test lead tips and check "0" is indicated on the display. - Connect the black test lead to the cathode side of the Diode and the red test lead to the anode side of the Diode.
Forward voltage of Diode is indicated on the display. - Connect the black test lead to the anode side of theDiode and the red test lead to the cathode side of the Diode. Normally, "OL" symbol is indicated on the display.
Conclusion: Diode is OK if the instrument complies with above items(3) and (4).
Note) Open-circuit voltage between measuring terminals is approx.1.5V.(measuring current approx. 0.4mA)
Capacitance Measurement
![]()
- To avoid the danger of getting electrical shock, never make measurement of the circuit in which electric potential exists.
- Do not make measurement when opening the instrument case.
- Make sure to discharge the capacitor before making measurement.
- Set the Function Selector Switch to "
" position. (Then, "AUTO" and "nF" symbols are indicated on the display.) - Press the SELECT key and "0" shall be indicated.(Then, "△" symbol is indicated on the display.)
- Connect the test leads to both ends of the resistanceunder test.
Measured value is indicated on the display.
Measuring unit "nF" / "uF" is automatically chosen and indicated due to the measured value.
Note) It may take some time according to the measuring capacitance.
Measuring capacitance < 4uF Measuring time is about 2seconds
Measuring capacitance < 40uF Measuring time is about 7seconds
Measuring capacitance < 100uF Measuring time is about 15seconds
Frequency Measurement
![]()
- To avoid the danger of getting electrical shocks, never make measurement on a circuit over 300V AC/DC. (electrical potential to ground 300V AC/DC)
- Do not operate Function Selector Switch during measurement
- Do not make measurement when opening the instrument case.
Frequency can be measured at ACV functions by pressing "Hz/DUTY" Switch. Concerning with the direction for use of "Hz/ DUTY" Switch, please reference to the item7-1 Hz/DUTY in this document.
Note) The minimum input can be measured is approx.1.5V.
How to use Function Switches
SELECT, Hz/DUTY Key
At each function, the actions of SELECT, Hz/DUTY key are different so please refer to following items and make active use of them.
- ACV function
(Above two keys act as Hz/DUTY switch key.) Capable of selecting the Voltage, Frequency or DUTY measurement mode.
At the initial condition, Voltage measurement has been selected for the ACV function.
By pressing "Hz/DUTY" key, measuring mode changes.
"Voltage" → "Frequency" → "DUTY" - DCV, Ω and Capacitance measurement
(Above two keys act as REL△ key.)
Indicate the difference between measured values. When any function ("DCV", "Ω" and "Capacitance") selected, the measured value can be stored by pressing SELECT key and after that, the difference between stored value and measuring value is indicated on the display
(△mark is keep lighting on the display while a value is stored)
The stored value can be released by pressing "SELECT" key again.
"Release" → "Memory" - Continuity check /Diode check function
(Change between Continuity check and Diode check) At the initial condition, "Continuity check" mode has been selected for the Continuity check / Diode check function.
By pressing "SELECT" key, measuring mode changes.
"Continuity check" → "Diode check"
The relative measurement is allowed in the following range.
*Measuring range = Full scale value at a range - initial value
DATA HOLD Key
The measured value can be hold at all functions.
By pressing "DATA HOLD" key, "DH" symbol indicated on the display and the indicated value can be held.
By pressing "DATA HOLD" key again, "DH" symbol disappears from the display and held data is released.
Auto Power Off
Auto power off function operates when about 15minutes passed after power on this instrument. When Auto power off function operates and the instrument powered off, the power-off statue returns to normal by pressing any key.
Battery Replacement
![]()
- Never open the instrument case when making measurement.
- To avoid getting electrical shock, be sure to remove test leads from the instrument when opening the instrument case in order to replace batteries.
- Remove the Portable holder from the instrument.
- Loosen one screw on the bottom of the instrumentand open the battery cover, then replace batteries. Battery: Two LR44(SR44)1.5V or equivalent
![KYORITSU - KEW 1018 - Battery Replacement Battery Replacement]()
Maintenance
Use a cloth dipped in water or neutral detergent for cleaning the instrument. Do not use abrasives or solvents.
Here you can download full pdf version of manual, it may contain additional safety instructions, warranty information, FCC rules, etc.
Need help?
Do you have a question about the KEW 1018 and is the answer not in the manual?
Questions and answers