● Save and keep the manual handy to enable quick reference whenever necessary. ● This instrument should only be used by suitably trained person and be sure to follow measurement procedures described in the manual. Kyoritsu assumes no responsibility for damage and injury caused by misuse or not following instructions in the manual.
Primary electrical circuit of the equipment, which is supplied power from a distribution board, and cable run from a distribution board to an outlet. CAT.Ⅲ: 2. Features This instrument, KEW1011, is a Digital Multimeter which designed for making measurement of the low-voltage equipment. (1) Designed to international safety standards (6) Auto-ranging function IEC 61010-1 Measurement CAT.Ⅲ...
Page 4
6.000V ±1.0%rdg±3dgt (50 / 60Hz) 60.00V Overload protection: 600V DC or AC ±1.2%rdg±3dgt (40∼400Hz) 0∼600V ACV rms Impedance: 10MΩ 600.0V (4 Auto ranging) 0.1V or less is outside the accuracy ±1.5%rdg±3dgt (50 / 60Hz) 600V ±1.7%rdg±3dgt (40∼400Hz) 600.0 μ A 0∼6000μA 6000 μ...
Page 5
Ranges Accuracy Test condition Note ±(5℃+5dgt)+accuracy of temperature probe −50∼0℃ ±(9° F+5dgt)+accuracy of temperature probe NiCr-NiSi sensor −58∼32° F ° C / ° F ±(3℃+2dgt)+accuracy of temperature probe If can not display bar graph on temperature 0∼150℃ 1℃/1° F ±(5° F+2dgt)+accuracy of temperature probe range 32∼302°...
Page 7
Measured value is indicated on the display. Note) When you make measurement of the voltage less than 0.1V at the AC6V range, the measurement value cannot be indicated correctly. Even if short-circuit the input line at the range of AC6V, 1∼3dgt may remain indicated. In that case, by pressing "REL△"...
Page 8
6-2-4 AC Current Measurement (up to 10A) CAUTION ● For safety sake, the measuring time on 10A range should be 15 seconds or less. Interval time between 2 measurements should be greater than 15 minutes. If you carry on making measurement continuously over 15 seconds, or make measurement in short interval time, it may cause error of measurement and instru- ment damage.
Page 9
Make sure that the "OL." symbol is indicated on the display at this bout, then short the test lead tips and check "0" is indicated on the display and check if the buzzer beeps. (4) Connect the test leads to both ends of the resistance under test. Measured value is indicated on the display.
7. How to use Function Switches 7-1 SELECT Key This key is to chose the measuring function at Ω/ Diode Check/ Continuity/ Capacitance function and Current function(uA, mA, A). Actions at each function are as below. ● Ω/ Diode Check/ Continuity / Capacitance function When setting the instrument to "Ω/ Diode Check/ Continuity/ Capacitance"...
8. Auto Power Off Auto power off function operates when about 15minutes passed after power on this instrument. When Auto power off function operates and the instrument powered off, the power-off statue returns to normal by pressing any key. And it is possible to release the Auto power off function. Turn the Function Selector Switch from OFF position to the function you want with pressing "SELECT"...
Page 12
1011 製造番号 保証期間 ご購入日 ( 年 月 日) より1カ年間 共立製品をお買い上げいただきありがとうございます。 保証期間内に通常のお取扱いで万一故障が生じた場合は、 保証規定により無償で修理いたします。 本書を添付の上ご依頼ください。 お名前 ご住所 〒 お電話番号 ( )−( )−( ) ◎保証規定をよくお読みください。 Kyoritsu reserves the rights to change specifications or Designs described in ◎本保証書は日本国内でのみ有効です。 this manual without notice and without obligations. ◎本保証書の再発行はいたしかねますので、大切に保管してくだ さい。 販売店名 92-1824C...