Utilities menu options

Utilities
Calibrate
Autolog
Comms
Test
30  Menus and Screens
Recalibrate
BFS must be
SunScan
connected by
cable
Restore
Factory
Calibration
Start
Date & time
Stop
Date & time
Read
hh:mm:ss
interval
Average
hh:mm:ss
interval
Ignore
night time
readings?
Start/Stop
logging
Sleep
Sets the individual sensors along the
SunScan wand to match the reading of
the connected BFS.
Restores the calibration of the SunScan
sensors that was determined at Delta-T.
SunData will operate the SunScan
unattended in whatever mode has been
setup, and will take regular periodic
readings. You may need to provide an
external power supply for the PDA if you
want to continue logging for many hours.
When you want to start logging. You can
set this a short time into the past, if you
want to synchronise readings to the hour
or day.
When you want to stop logging. Must be in
the future!
How often you want to take readings
Minimum 00:00:01, maximum 23:59:59.
How often you want to average your
readings. Should be a multiple of the read
interval. Set to zero for no averaging.
If checked, SunData will stop taking
readings during the night, and start again
at dawn.
SunData will start taking readings
according to the Autolog settings. The
display will show a record of readings
taken recently, and the time the next
reading is due.
If logging is in progress, this will stop
logging, and return to the Autolog setup
screen.
Puts the PDA into sleep mode. It will wake
up again briefly every time a reading is
taken.
This screen gives direct access to the
SunScan via the serial port. It should only
be used under instruction from Delta-T.