PhysioGo.Lite Electro
User manual

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the PhysioGo.Lite Electro and is the answer not in the manual?

Questions and answers

Summary of Contents for Astar PhysioGo.Lite Electro

  • Page 1 PhysioGo.Lite Electro – User manual...
  • Page 3: Table Of Contents

    1. Basic information and contents Contents INTRODUCTION ............................... 6 ................................6 ANUFACTURER .............................. 6 ISK MANAGEMENT PROCESS INTENDED USE ................................. 7 ................................8 NTENDED USERS ................................8 SER TRAINING WARRANTY AND MANUFACTURER'S RESPONSIBILITY....................9 OPERATIONAL SAFETY ............................10 ..........................10 AINS SUPPLY AND OPERATION MODE ......................
  • Page 4 ............................82 ABLE CONDITION CONTROL ........................83 ERIFICATION OF ELECTRODES CONDITION ..............................84 PECIAL MESSAGES ............................... 84 TEST PROCEDURE ..............................85 ROUBLESHOOTING ..............................86 USE REPLACEMENT SPECIFICATION AND ACCESSORIES ......................... 87 User Manual – PhysioGo.Lite Electro page 4 / 96...
  • Page 5 ..............................87 ECHNICAL DATA ..............................89 PARAMETERS ............................. 91 TANDARD ACCESSORIES ............................. 91 PTIONAL ACCESSORIES APPENDIX A. SYMBOL DESCRIPTION, I(T) CURVE DIAGRAM ................... 92 ........................92 ONTROLLER ACCESSORIES PACKAGING – ........................94 WITCHED MODE POWER SUPPLIES CASING Table of figures Figure 5.1 General view ................................
  • Page 6: Introduction

    Read this manual carefully before starting the unit operation! Follow the recommendations presented in this manual! The electrotherapy unit PhysioGo.Lite Electro should be installed by the seller. The recipient has the right to insist on the product operation training. The unit may only be operated by qualified personnel or under supervision of such personnel! WARNING: The device is intended for adult patients only.
  • Page 7: Intended Use

    2. Intended use 2. Intended use Electrotherapy unit PhysioGo.Lite Electro is an active, non-invasive therapeutic device, intended for carrying our treatment procedures using: bipolar (bidirectional) and unipolar (unidirectional) low frequency currents, • • bipolar (bidirectional) and unipolar (unidirectional) medium frequency currents.
  • Page 8: Intended Users

    User training The PhysioGo.Lite Electro user has to be properly trained in the device safe and effective use, before starting the operation. Training in the rules of operation can be carried out by representatives of the manufacturer or seller, based on this user manual.
  • Page 9: Warranty And Manufacturer's Responsibility

    3. Warranty and manufacturer's responsibility 3. Warranty and manufacturer's responsibility The manufacturer warrants the controller to be free of faults for the period of time and conditions stated in Warranty Certificate. The manufacturer also provides post-warranty service for a period of 10 years from launching the unit onto the market.
  • Page 10: Operational Safety

    LED indicator located on the housing of the switched-mode power supply type GSM60B24-P1J by Mean Well. Recommendations related to isolation the device from the supply mains: do not position the PhysioGo.Lite Electro so that it is difficult to operate the disconnection of the device • from the supply mains, to isolate the device from the supply mains, hold the mains socket-outlet with one hand, grasp the mains •...
  • Page 11: Warnings And Safety Notes

    WARNINGS and safety notes The PhysioGo.Lite Electro unit has been designed and manufactured in such a way that its use does not cause a hazard to the health and safety of patients, operators and third parties, as well as the unit provides therapeutic benefits to patients if it is operated in appropriate conditions and in accordance with its intended purpose.
  • Page 12 Special caution must be kept during electrotherapy treatments in children and older people. • Battery use (optional): Rechargeable battery UR18650A 5S1P is designed for use only with Astar PhysioGo.Lite range of devices. • • In case of mechanical damage to the battery module, there is a risk of fire, explosion or burns, due to used lithium-ion cells.
  • Page 13: Explosion Proof Environment

    WARNING: Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be used no closer than 30 cm (12 inches) to any part of the PhysioGo.Lite Electro, including cables specified by the manufacturer. Otherwise, degradation of the performance of this equipment could result.
  • Page 14: Essential Performance

    4. Operational safety Essential performance For the PhysioGo.Lite Electro device, the essential performance means the generation of current and voltage signals with frequencies, shapes and amplitudes corresponding to the waveforms recognized and used in this therapy. Unidirectional (unipolar) and / or bidirectional (bipolar) currents are available.
  • Page 15: Disposal

    4. Operational safety Disposal In case, when the disposal of the unit will become necessary (e.g. after elapse of its service life), please contact the manufacturer or manufacturer representative, which must react in an appropriate way i.e. collecting the unit from the user.
  • Page 16: Unit Description

    5. Unit description General characteristics Electrotherapy unit PhysioGo.Lite Electro is a highly specialized medical device based on modern microprocessor platform. The unit has a plastic console casing. It is equipped with a color touch LCD display with a diagonal of 12,7 cm (5’’).
  • Page 17: Front Panel

    5. Unit description Front panel Arrangement of front panel components is shown in Figure 5.3. Figure 5.3 Arrangement of front panel components Table 5.1 Description of front panel components Symbol Description Function The unit is equipped with a touch-sensitive display. The screen clearly Liquid crystal display displays all information related to the device operation.
  • Page 18: Operation Status And Battery Level Indicators

    Yellow Battery error. Turn the unit off using STANDBY key and 5 pulses switch the power switch off. Restart it after 10 seconds. If the problem repeats, contact your authorized service. User Manual – PhysioGo.Lite Electro page 18 / 96...
  • Page 19: Battery Installation

    5. Unit description Battery installation PhysioGo.Lite Electro can be optionally fitted with a battery. Users can install the battery by themselves. WARNING: Before performing the below actions, turn the device power off and disconnect the SMPS from the mains. The battery assembly method is illustrated in the following figures.
  • Page 20: Current And Voltage Stabilization - Cc And Cv Mode

    The calibration settings for CC mode are being defined with the load which has a resistance of 500 Ω. The calibration settings for CV mode are being entered in the idle operation mode of the unit. User Manual – PhysioGo.Lite Electro page 20 / 96...
  • Page 21: Protection

    5. Unit description Protection 5.6.1 Detection of a high resistance in the patient circuit In the case, when at the start or during an electrotherapy / combined therapy treatment procedure, the device detects a state of high resistance in the patient circuit, which may be caused by: •...
  • Page 22: Electrodes Condition Control

    (e.g. training programs, modulations, pulse current). Detailed information and symbols are provided – see Table 5.6. User Manual – PhysioGo.Lite Electro page 22 / 96...
  • Page 23 5. Unit description Table 5.6. Currents classification Signal Shape of the Classification Threshold for current warning TENS pulse currents symmetric 10 mA asymmetric bipolar 10 mA alternating 10 mA Isoplanar interferential current sinusoidal bipolar 10 mA Interferential current with vector sinusoidal bipolar 10 mA...
  • Page 24: Device Installation And Start-Up

    It is recommended to operate with the unipolar currents using metal – tin or aluminum electrodes, as they wear out much slower than the electrodes made from other materials. User Manual – PhysioGo.Lite Electro page 24 / 96...
  • Page 25: First Operation

    6. Device installation and start-up As optional accessories you can purchase self-adhesive electrodes in different dimensions. This type of electrodes is suitable for use with bipolar currents, especially TENS currents. They shall not be used for therapy with unipolar currents! Selection of the electrodes type to a particular treatment should be based on doctor’s or physiotherapist’s knowledge and experience.
  • Page 26: Language

    • 6.3.4 Functional settings 6.3.4.1 Channel operation mode selection This function allows the user to set preferred style of unit operation. User Manual – PhysioGo.Lite Electro page 26 / 96...
  • Page 27: Control Functions

    6. Device installation and start-up Option Explanation Manual mode – automatically After therapy selection, the unit is set in manual mode of the operation. Program mode – automatically After therapy selection, the unit is set in program mode of the operation. After therapy selection, the unit displays a window with a list of operation Mode selection pop-up mode selection options.
  • Page 28: Information

    The „Show logs” button supports service diagnostics, by displaying information about all saved device errors. 6.3.6.5 Unit statistics Provides information about the number of treatment procedures performed. Statistics can be deleted. If you want to delete statistics, press Delete programs counters button. User Manual – PhysioGo.Lite Electro page 28 / 96...
  • Page 29: Transport Position - Trolley For The Unit

    6. Device installation and start-up Transport position – trolley for the unit Step Description Click the UNIT ON / OFF (STANDBY) button on the front panel A message will appear on the display: Press to shut down the system correctly. Disconnect from the controller the power cord and accessories.
  • Page 30: Unit Operation

    (e.g. aloe Vera) if there are no viscose or gauze pads. Properly attach the electrodes with viscose pads to the patient’s body, e.g. by Velcro belts, elastic • bandage or sand bags. User Manual – PhysioGo.Lite Electro page 30 / 96...
  • Page 31: Screen Configuration

    7. Unit operation In the place of the arrangement of the electrodes it is necessary to estimate the continuity of the skin • and exteroceptive sensation. • In order to decrease the resistance of skin you can clean it with alcohol or water with soap, after wiping leave the skin moist.
  • Page 32: General Configuration

    • The color of the tab is identical to the background color of the edition field. In the background, other operating tabs are black. User Manual – PhysioGo.Lite Electro page 32 / 96...
  • Page 33: Current Selection Screens

    7. Unit operation The table below explains the meaning of the symbols presented in tabs: Symbol Description Channel 1 Selected single circuit electrotherapy – socket A Channel 1 Selected dual circuit electrotherapy – socket A and B Channel 2 Selected single circuit electrotherapy – socket B Ongoing treatment (white symbol) Treatment interrupted (white symbol) Error (yellow symbol)
  • Page 34: Limitations

    Figure 7.2 Screenshot sample view for dual circuit electrotherapy A+B Symbol Description current name and shortened information on treatment Manual mode parameters Program mode program name A circuit identification and amplitude value B circuit identification and amplitude value User Manual – PhysioGo.Lite Electro page 34 / 96...
  • Page 35: Operation With Preset Treatment Programs And Sequences

    7. Unit operation Symbol Description Indication of a current flow with an amplitude above a certain threshold (see 5.6.3) Presentation of the treatment elapsing time Current identification (AC/DC) (see 5.6.5) Tab field – channel 1 Figure 7.3 Screenshot sample view for single circuit electrotherapy A and B Figure 7.4 Screenshot sample view for electrotherapy sequences Symbol Description...
  • Page 36: Figure 7.5 Information Screen Sample View

    Select the program / sequence from the list. Prepare the patient for the treatment according to indications in point 7.1 Press the key Using the keys set the current or voltage amplitude. User Manual – PhysioGo.Lite Electro page 36 / 96...
  • Page 37: Favorite Programs

    7. Unit operation Favorite programs The function offers quick access to frequently used preset programs and sequences without browsing the entire list. To add or remove the program from the favorite list, follow the instructions: Step Description Prepare the unit to work with preset treatment programs and sequences (see section 7.4). Select program / sequence.
  • Page 38: Manual Mode Operation

    Correct the parameters. Press the button from main menu. Select the item number under which the program will be saved. Confirm your choice with Enter or edit the program name. Press the button User Manual – PhysioGo.Lite Electro page 38 / 96...
  • Page 39: User Sequences

    7. Unit operation Removal of user program: Step Description Prepare the unit to work in the program mode (see section 7.5). Select the option User programs from Program modes menu. Confirm your choice by pressing the selected field again. Select the program which will be deleted. Press the button from main menu.
  • Page 40 Use the sequence edition tools described below to make changes in the created sequence. Press the button to save the sequence. Enter the name. Then press Press the button to escape from the sequence editor. User Manual – PhysioGo.Lite Electro page 40 / 96...
  • Page 41 7. Unit operation User sequence edition: Step Description Prepare the unit to work in the program mode (see section 7.5). Select the option User sequences from Program modes menu. Confirm by pressing the selected field once again. Press the button from main menu.
  • Page 42: I/T Curve

    Press to see again the graph. Press to save the curve. Enter the name and press Press the key to escape from the I/t curve mode. User Manual – PhysioGo.Lite Electro page 42 / 96...
  • Page 43: Safe Shutdown Procedure

    7. Unit operation It may happen that the muscle reaction is not observed for triangular pulses of 1 s pulse duration. Then you should interrupt the procedure and start it again from 500 ms pulse duration. For the time value of 1 s should be set 0 mA value.
  • Page 44: Definitions And Parameters

    The programs which feature relatively long rest time in relation to contraction phase give appropriate conditions for muscle fiber regeneration. The programs which have decisively shorter rest phase shall only be applied to electro-gymnastics with healthy persons or to cause tonolitic effect. User Manual – PhysioGo.Lite Electro page 44 / 96...
  • Page 45: Treatment Timer

    8. Definitions and parameters Treatment timer The treatment time is adjustable from 1 - 60 minutes with a step of 1 minute. It is also possible to adjust manually by "holding" the field with time by means of a pen / finger – in increments of 1 s (Figure 8.2) Figure 8.1.
  • Page 46 7 pulse sequences, 4 Hz BURST mode FM = BURST 9 pulse sequences, 2 Hz 9 pulse sequences, 4 Hz Amplifier CC – stabilized output current operation mode CV – stabilized output voltage User Manual – PhysioGo.Lite Electro page 46 / 96...
  • Page 47: Interferential Currents

    8. Definitions and parameters Symbol on the Symbol Description treatment Available parameters screen 1 – 140 mA in CC mode Regulation step: Maximum • 0,5 mA in the range of 1-140 mA amplitude 2 – 140 V in CV mode, max. 140 mA Regulation step: •...
  • Page 48: Kotz' Current (Russian Stimulation)

    Current of 2500 Hz frequency is joined to form rectangular trains or ”bursts” with length equal to the pause time (e.g. pulse current with 20 Hz frequency consists of bursts lasting 25 ms and pauses of equal length). User Manual – PhysioGo.Lite Electro page 48 / 96...
  • Page 49 8. Definitions and parameters The bipolar pulse current of such duty cycle is subjected to amplitude modulation to obtain smooth increase and reduction of output current within patient circuit (by the use of a training program), which results in mild muscle contraction and relaxation effect with determined activity and rest phase.
  • Page 50: Medium Frequency Currents

    Rise time 1 s Contraction phase 2 s Fall time 1 s Training Rest phase 2 s program Rise time 1 s Contraction phase 3 s Fall time 1 s Rest phase 8 s User Manual – PhysioGo.Lite Electro page 50 / 96...
  • Page 51 8. Definitions and parameters Symbol on the Symbol Description treatment Available parameters screen Rise time 1 s Contraction phase 5 s Fall time 1 s Rest phase 5 s Rise time 1 s Contraction phase 5 s Fall time 1 s Rest phase 10 s Rise time 2 s Contraction phase 10 s...
  • Page 52: Sp-Tens Pulse Current

    Rise time 1 s Stimulation phase 4 s Fall time 1 s Rest phase 6 s Rise time 1 s Stimulation phase 4 s Fall time 1 s Rest phase 10 s User Manual – PhysioGo.Lite Electro page 52 / 96...
  • Page 53: Diadynamic Currents

    8. Definitions and parameters Symbol on Symbol Description the treatment Available parameters screen Rise time 2 s Stimulation phase 6 s Fall time 2 s Rest phase 10 s Amplifier operation CC – stabilized output current mode CV – stabilized output voltage 1 –...
  • Page 54: Galvanic Current

    0,3 – 80 mA in CC mode Regulation step: Maximum amplitude • 0,1 mA in the range of 0,3-10 mA 0,5 mA in the range of 10-80 mA • User Manual – PhysioGo.Lite Electro page 54 / 96...
  • Page 55: Ultra Reiz Current (Träbert ' S Current )

    8. Definitions and parameters Ultra Reiz current (Träbert’s current) The Ultra Reiz current (Träbert’s current) is a special case of rectangular pulse current. This is a current with rectangular shape, pulse duration 2 ms and pause duration 5 ms. These settings are not adjustable. Ultra Reiz current is used in segmentary and local applications.
  • Page 56: Unipolar And Bipolar Sine Surge

    Amplitude rise time 3 s program or AM / PT Hold time of maximum amplitude 3 s training Amplitude fall time 3 s program Amplitude rise time 6 s Amplitude fall time 6 s User Manual – PhysioGo.Lite Electro page 56 / 96...
  • Page 57 8. Definitions and parameters Symbol on the Symbol Description treatment Available parameters screen Amplitude rise time 6 s Hold time of maximum amplitude 12 s Amplitude fall time 6 s Rise time 1 s Contraction phase 2 s Fall time 1 s Rest phase 2 s Rise time 1 s Contraction phase 3 s...
  • Page 58: Leduc's Current

    Basic pulse Constant, 100 Hz frequency Training No training program, non-editable program parameter For such polarization setting red plug is a Polarization positive electrode, and black plug is a negative electrode. User Manual – PhysioGo.Lite Electro page 58 / 96...
  • Page 59: Rectangular Pulse Currents

    8. Definitions and parameters Symbol on Symbol Description the treatment Available parameters screen For such polarization setting red plug is a negative electrode and black plug is a positive electrode. Warning means that such a setting is the reverse of commonly accepted way of marking polarity.
  • Page 60 Rise time 3 s Contraction phase 15 s Fall time 3 s Rest phase 30 s Rise time 3 s Contraction phase 15 s Fall time 3 s Rest phase 50 s User Manual – PhysioGo.Lite Electro page 60 / 96...
  • Page 61: Triangular Pulse Currents

    8. Definitions and parameters Symbol on Symbol Description the treatment Available parameters screen For such polarization setting red plug is a positive electrode, and black plug is a negative electrode. For such polarization setting red plug is a negative electrode and black plug is a Polarization positive electrode.
  • Page 62 0,5 mA in the range of 10-140 mA • current 1,5 – 100 V in CV mode, max. 80 mA Regulation step: • 0,5 V in the range of 1,5-100 V User Manual – PhysioGo.Lite Electro page 62 / 96...
  • Page 63: Neofaradic Pulse Currents

    8. Definitions and parameters Neofaradic pulse currents The neofaradic current is a low frequency pulse current, which is a modern version of the faradic current. It consists of positive rectangular or triangular pulses with a duration of 1 ms and a variable frequency (and therefore a variable pause time).
  • Page 64: Microcurrents

    Frequency rise time 6 s Frequency fall time 6 s Frequency rise time 6 s Hold time of maximum frequency 6 s Frequency fall time 6 s Hold time of basic frequency 6 s User Manual – PhysioGo.Lite Electro page 64 / 96...
  • Page 65: Ig Pulses

    8. Definitions and parameters Symbol on Symbol Description the treatment Available parameters screen Frequency rise time 12 s Frequency fall time 12 s 0 – 1000 µA in CC mode Maximum Regulation step: amplitude • 50 µA within the range IG pulses IG pulses are triangular unipolar or bipolar pulse currents.
  • Page 66: Ems

    Its aim is to maintain the contraction ability of the properly innervated muscle, to prevent muscle atrophy and to restore muscle strength and mass. User Manual – PhysioGo.Lite Electro page 66 / 96...
  • Page 67 8. Definitions and parameters Parameters description: Symbol on the Symbol Description treatment Available parameters screen Treatment time 1 – 60 minutes, 1 minute step Shape of the Symmetric current Possible settings: 25 µs, 50 µs, 75 µs, 100 µs, 150 µs, 200 Pulse duration t_imp µs, 250 µs, 300 µs, 400 µs, 500 µs...
  • Page 68: H-Waves

    Low frequency increases the lymph and blood flow, accelerates toxin outflow and reduces swelling. In the higher frequency mode, the H-waves work by their influence on the sodium pump function within the nerve and thus produce a profound anaesthetic / analgesic effect. User Manual – PhysioGo.Lite Electro page 68 / 96...
  • Page 69: Exponential Pulses

    8. Definitions and parameters Parameters description: Symbol on the Symbol Definition treatment Available parameters screen Treatment time 1 – 60 minutes, 1 minute step Shape of the Symmetric current Pulse duration t_imp Constant, 11,2 ms Non-editable parameter, calculated from the formula Pause duration t_pause (1/f_bas)-11,2...
  • Page 70 17,1 Hz 10 Hz 9,2 Hz 5 Hz 8,6 Hz 5 Hz 4,6 Hz 3 Hz 2,8 Hz 1 Hz 1,5 Hz 1 Hz 1,7 Hz 1 Hz 0,9 Hz 0,5 Hz User Manual – PhysioGo.Lite Electro page 70 / 96...
  • Page 71: Hufschmidt Stimulation

    8. Definitions and parameters Hufschmidt stimulation Hufschmidt stimulation is a method of two-channel alternating electrostimulation of spastic and antagonistic muscles. In the therapy, a single rectangular pulse is applied to the spastic muscle, then the antagonistic muscle is stimulated with a single rectangular pulse (with a certain delay, during rest of spastic muscle). Parameters description: Symbol on the Symbol...
  • Page 72: Tonolysis

    TRI-UNI • 40 kHz modulated by rectangular (channel B) pulses with 4 kHz frequency • Duty factor 50% Triangular bipolar Frequency and current shape: TRI-BI • Rectangular 4 kHz User Manual – PhysioGo.Lite Electro page 72 / 96...
  • Page 73 8. Definitions and parameters Symbol on Symbol Description the treatment Available parameters screen Sinusoidal unipolar Frequency and current shape: SIN-UNI • 40 kHz modulated by rectangular pulses with 4 kHz frequency • Duty factor 50% Pocket frequency f_trig Regulation in the range of 0,2 Hz – 10 Hz Time lag between t_del Regulation in the range of 5 ms –...
  • Page 74 0,5 V in the range of 1,5-100 V • 1,5 – 100 V, max. 80 mA for unipolar stimulating pulses (channel B) Regulation step: • 0,5 V in the range of 1,5-100 V User Manual – PhysioGo.Lite Electro page 74 / 96...
  • Page 75: Indications And Contraindications

    9. Indications and contraindications 9. Indications and contraindications Indications 9.1.1 TENS and SP-TENS current biological impact: pain relief, improvement of blood circulation, stimulation of muscles, stimulation of nerve fibers with varying effect depending on frequency range, pulse width and modulation type: •...
  • Page 76: Kotz' Current - Russian Stimulation

    • • chilblains • oedemas, subcutaneous and intramuscular extravasations emphysema, subcutaneous emphysema • 9.1.6 Galvanic current biological impact: • dilatation of peripheral vessels ion movement within tissues • User Manual – PhysioGo.Lite Electro page 76 / 96...
  • Page 77: Ultra Reiz Current

    9. Indications and contraindications therapeutic application: drug administration (iontophoresis) • • circulation disorder • diagnostics of internal inflammatory focuses (galvano-palpation) 9.1.7 Ultra Reiz current biological impact: • reduction of increased activity of the sympathetic system decrease in muscle tone • pain relief •...
  • Page 78: Triangular Impulses

    • • ulceration aesthetic medicine • 9.1.14 IG pulses biological impact: • pain relief decrease in muscle tone • therapeutic application: neuralgias • • muscle pains joints degenerative changes • User Manual – PhysioGo.Lite Electro page 78 / 96...
  • Page 79: Ems

    9. Indications and contraindications • post-traumatic conditions after joints and muscles injuries constipation • • peripheral circulation disorders 9.1.15 EMS biological impact: activation of muscles through unimpaired peripheral nerves to increase contraction force and sensory • consciousness obtaining correct movement patterns •...
  • Page 80: Contraindications For Electrotherapy

    • • cases, when the skin cannot be moistened recent surgery, unstable fracture, osteoporosis • • epilepsy exception: TENS applied on limbs can reduce epileptic activity  User Manual – PhysioGo.Lite Electro page 80 / 96...
  • Page 81: Maintenance, Cleaning, Disinfection

    10. Maintenance, cleaning, disinfection 10. Maintenance, cleaning, disinfection NOTE: The warranty does not cover any damage due to a failure to adhere to the recommendations stated in this chapter. NOTE: Before attempting to perform following operations isolate the unit and switch mode power supply from the mains supply! The activities of maintenance, cleaning and disinfection of device components should be realized at: •...
  • Page 82: Cleaning And Disinfection Of The Electrotherapy Accessories

    Additionally, you should move the cable near the plugs. Observe the indication shown on the display. Press the button Run the test Evaluate the usage of cables according to instructions below: Cable in good condition User Manual – PhysioGo.Lite Electro page 82 / 96...
  • Page 83: Verification Of Electrodes Condition

    10. Maintenance, cleaning, disinfection Step Description Reducing the signal level indicates the cable damage . To leave Setup mode, press the key To escape the test mode press the key Alternative method: cable plug should be inserted into the output socket, and plugs from the side of electrodes should be short-circuited.
  • Page 84: Special Messages

    Warnings Self-test procedure Each time the PhysioGo.Lite Electro unit is started, a special self-test procedure is performed during which all modules and functional blocks of the device are tested. If any errors or damage are detected, the relevant information will be presented on the display screen. The module where the irregularity has been detected is blocked and the associated function is not available.
  • Page 85: Troubleshooting

    10. Maintenance, cleaning, disinfection In this situation, you must disconnect the unit from the mains and contact an authorized service for inspection and a possible repair. Table 10.2. The "hardware" error coding system Error code Error description SDRAM self-test error No communication with the SD card No communication with the TSC controller in the LCD Program defect in the processor FLASH memory (CRC)
  • Page 86: Fuse Replacement

    Connect the power supply cord to the device socket. Then connect the mains cable to the mains socket. Switch the power switch on and start the device. Check the device operation. User Manual – PhysioGo.Lite Electro page 86 / 96...
  • Page 87: Specification And Accessories

    Treatment timer: Ranges and resolutions: defined in chapter 8 Time accuracy: General: Mains supply: 100-240 V; 50/60 Hz PhysioGo.Lite Electro controller supply: 24VDC; 2,5A Mains fuses: size 5x20mm, T3,15L250V; 3,15 A, 250 V Type of memory backup battery: CR2032 Unit weight: max.
  • Page 88 Operation conditions: Temperature range: +15÷+30°C Relative humidity: 30÷75% Pressure range: 700÷1060 hPa (70 – 106 kPa) Transport conditions: Temperature range: -10÷+45°C Relative humidity: 20÷95% Pressure range: 700÷1060 hPa (70 – 106 kPa) User Manual – PhysioGo.Lite Electro page 88 / 96...
  • Page 89: Emc Parameters

    RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the PhysioGo.Lite Electro unit is used exceeds the applicable RF compliance level above, the unit should be observed to verify normal operation.
  • Page 90 Voltage interruptions 250 cycles for 50 Hz Complies IEC 61000-4-11 300 cycles for 60 Hz Immunity test Compliance level Proximity fields from RF wireless communications equipment according to 8.10 IEC Complies 60601-1-2:2014 User Manual – PhysioGo.Lite Electro page 90 / 96...
  • Page 91: Standard Accessories

    11. Specification and accessories Standard accessories Name Quantity PhysioGo.Lite Electro controller A-UE-AST-PLE Switch mode power supply – type HPU63B-108 by Sinpro or GSM60B24-P1J by Mean Well Mains cable Patient’s cable: a) Channel A a) A-AE-AST-KPPL2M_A b) Channel B b) A-AE-AST-KPPL2M_B...
  • Page 92: Appendix A. Symbol Description, I(T) Curve Diagram

    Degree of protection provided by enclosures (IP code), based on IEC 60529 Fuse, symbol IEC 60417-5016 Unit version Serial number, symbol ISO 7000-2498 Batch code, symbol ISO 7000-2492 Catalogue number, symbol ISO 7000-2493 Medical device, symbol 5.7.7. of ISO 15223-1:2020 standard User Manual – PhysioGo.Lite Electro page 92 / 96...
  • Page 93 Appendix A Symbol Definition Unique Device Identifier, symbol 5.7.10. of ISO 15223-1:2020 standard Disposal of used devices together with other waste is prohibited, complied with the requirements of WEEE General symbol for recovery/recyclable, symbol ISO 7000-1135 Operator's manual; operating instructions, symbol ISO7000-1641 Switch mode power supply socket, direct current, symbol IEC 60417-5031 Switched-mode power supplies identification Follow operating instructions, symbol ISO 7010-M002...
  • Page 94: Switched-Mode Power Supplies - Casing

    Eurasian Customs Union. GSM60B24-P1J Caution, symbol ISO 7000-0434A (Mean Well) GSM60B24-P1J Dangerous voltage, symbol IEC 60417-5036 (Mean Well) For indoor use only, symbol IEC 60417-5957 Class II equipment, symbol IEC 60417-5172 User Manual – PhysioGo.Lite Electro page 94 / 96...
  • Page 95 Appendix A Symbol Description SMPS type Compliance with the RoHS directive SJ/T 11364-2014 (China). The number GSM60B24-P1J indicates the service life of an environmentally friendly electric and electronic (Mean Well) product. Disposal of used devices together with other waste is prohibited, complied with the requirements of WEEE GSM60B24-P1J Upper limit of temperature, symbol ISO 7000-0533...
  • Page 96 Appendix A User Manual – PhysioGo.Lite Electro page 96 / 96...