Status tabs ..............................25 7.2.3 Limitations ..............................26 ..............................26 ISPLAY DESCRIPTION ................... 27 PERATION WITH PRESET TREATMENT PROGRAMS AND SEQUENCES ................................... 28 AVORITES ............................. 29 ANUAL MODE OPERATION ................................29 SER PROGRAMS Etius – User Manual Page 3 / 60...
Page 4
Figure 5.2. View of the left and right panel of the unit ......................14 Figure 5.3. Arrangement of keyboard components ......................... 15 Figure 5.4. Rough diagram of the output circuit of the Etius unit working in CV mode (one channel)........17 Figure 6.1. Electrotherapy sockets ............................20 Figure 6.2.
Read this Guide carefully before starting the unit operation! Follow the recommendations presented in this Guide! Electrotherapy therapy unit Etius should be installed by the seller. The recipient has the right to insist on the product operation training. The unit may only be operated by qualified personnel or under supervision of such personnel! WARNING: The device is intended for adult patients only.
(bidirectional) medium frequency currents and unipolar (unidirectional) medium frequency currents • modulated by low frequency waveforms. Etius unit is equipped with two fully independent electrotherapy treatment channels with the possibility of operation on 4 electrodes treatment procedures. Detailed information about the available configuration is presented in chapter 7.
2.2 User training The Etius user has to be properly trained in the device safe and effective use, before starting the operation. Training in the rules of operation can be carried out by representatives of the manufacturer or seller, based on this user manual.
On demand, the producer makes available technical diagrams, parts lists, descriptions, instructions for calibration or other helpful information to appropriately qualified user's technical staff to repair these parts of unit, which are described by the producer as a reparable. Etius – User Manual Page 8 / 60...
4.1 Mains supply and operation mode The Etius unit is designed for supply from AC mains with rating 230 V ± 10%, 50/60Hz. It is a medical device under safety class I, type BF. The unit may be used only in rooms, where the electric system is executed in compliance with standards in force.
Page 10
• Do not position Etius so that it is difficult to operate the disconnection of the device from the supply mains. Do not remove warning signs and labels put by the manufacturer on the unit casing and casings of •...
Page 11
4. Operational safety Electrodes should not be placed alongside the carotid artery (carotid sinus). • • It is prohibited to leave the patients unattended during treatments of electrotherapy. It is necessary to take precautions in case of the occurrence of surface metal implants in the spot of •...
Etius unit. The unit meets requirements of electromagnetic interference emission and immunity standards and shall not pose a threat to correct operation of the other devices.
4. Operational safety 4.6 Technical supervision The User of Etius unit should perform technical inspection of the unit at year's intervals. The inspection should be performed by a unit authorized by the manufacturer. The inspection is performed at the user's expense.
7.2. Etius features console casing with top made of light-grey plastic. All indicators are located on the upper surface. The mains switch, fuse socket are mains socket are located on the left side of casing. Output sockets for connecting patient’s cables and patient's stop switch are located on the right side of casing.
5.Unit description 5.2 Keyboard Arrangement of keyboard components is shown in figure 5.3. Figure 5.3. Arrangement of keyboard components Description Functionality Liquid crystal The unit is equipped with a LCD with backlight and a resolution of 240x64 lines. display The screen clearly displays all information related to the device operation. Spotlight indicators inform about current flow in a given channel.
On the left side panel are located: unit mains switch, • • fuse socket, • mains socket. On the right side panel are located: • socket for connecting patient's cable, • patient's stop switch socket. Etius – User Manual Page 16 / 60...
Kirchhoff voltage law) between the unit’s output resistance and resistance of a load. Rough diagram of operation system in CV mode for one channel is presented in figure 5.4. Figure 5.4. Rough diagram of the output circuit of the Etius unit working in CV mode (one channel). Symbols used in figure: E –...
In the case when the difference between the setting and output value is higher than 20%, the treatment will be interrupted and the message "Current limit exceeded" will appear on the display. Etius – User Manual Page 18 / 60...
5.Unit description 5.7.4 Overcurrent in CV mode While performing the electrotherapy treatment in stabilized output voltage mode (CV), the unit controls whether the maximum current value is not beyond the limits. If the voltage setting for current value is higher than the allowed limit (specified for the stabilized output current in CC mode), the treatment will be interrupted and the message "Overcurrent"...
Cables are terminated with banana type 4 mm or 2 mm plugs – two are red and the other two are black. Channels are marketed with appropriate symbols. Electrodes should be connected to those plugs. Figure 6.2. Connection of electrodes Etius – User Manual Page 20 / 60...
6.1.2 Connection in combined therapy It is possible to apply Etius to combined therapy in connection with ultrasound device, for example Sonaris. To perform a combined therapy treatment session connect red plug to the ultrasound unit and black plug to the electrode.
Approval of selection is performed by pressing the key. Escape without approval (settings from before modification remain) is performed by pressing the key. Etius – User Manual Page 22 / 60...
6.Device installation and start-up 6.2.5 Removal of user-defined programs Selection of this option enables overall deletion of programs saved by the user. In order to delete all user-defined programs, follow the instructions shown on the display. While editing settings, it is unnecessary to repeat deletion of user-defined programs several times.
Apply the gel (e.g. aloe vera) coupling the electrodes with the patient’s body if there are no • viscose or gauze pads. Properly attach the electrodes with viscose pads to the patient’s body, e.g. by velcro belts, elastic • bandage or sand bags. Etius – User Manual Page 24 / 60...
7. Unit operation In the place of the arrangement of the electrodes it is necessary to estimate the continuity of the skin • and exteroceptive sensation. • In order to decrease the resistance of skin you can clean it with alcohol or water with soap, after wiping leave the skin moist.
Electrotherapy circuit – socket A Circuit field Electrotherapy circuit – socket B Amplitude settings: Amplitude field • current and voltage for electrotherapy circuits Etius – User Manual Page 26 / 60...
The list of all disease entities with parameters is included in the guide "Preset Treatment Programs and Sequences - Etius Family and Polaris 2" which is attached to the unit's accessories. The values of the preset treatment programs parameters are based on the available literature data and they are determined as average values.
Exclusion is possible only from the favorites list or a list of preset treatment programs level. If for a given therapy no program has been selected as a “favorite”, after going over the favorites, the message "Favorites list is empty" will be shown. Etius – User Manual Page 28 / 60...
7. Unit operation 7.6 Manual mode operation Symbol definition and parameters range are given in chapter 8. Step Description Switch on the unit. Choose the treatment channel 1 or 2. Select the therapy according to the point 7.2.1 If the unit is in program mode, press the key.
7.8 I/t curve It is recommended to use the results obtained by determining I/t curve to create treatment programs that will be executed exclusively with the Etius, Etius U, Etius LM and Etius ULM units. Step Description Switch on the unit.
7. Unit operation 7.9 Safe shutdown procedure The work flow for the safe termination of the operation: Step Description Proceed with the treatment to the end or stop it by pressing the key. Put the equipment back. Turn the device off using the mains switch located on the left side of the device. If you do not use the device for a long time, remove the mains cable plug from the mains socket.
Parameters description: Symbol Description Available parameters Treatment time 1 - 60 minutes continuous Continuous or interrupted shape pulse frequency is 4kHz of the current duty factor 80% Etius – User Manual Page 32 / 60...
• 8.1.3 Diadynamic currents Diadynamic currents generated by the Etius unit are medium frequency currents (carrier wave is 40 kHz) modulated by low frequency current, which comply with traditional diadynamic currents described by Bernard. There are no significant therapeutic differences between such an approach to diadynamic currents generation and traditional method.
8.Definition and parameters 8.1.4 Rectangular and triangular pulse currents Pulse currents generated in the Etius unit are medium frequency unipolar currents modulated by low frequency rectangular or triangular current. Rectangular pulse current consists of pulse sequence of rectangular shape and independently adjusted times of pulse and pause.
Page 35
8.Definition and parameters Symbol Description Available parameters Rise time 1 s Contraction phase 5 s Fall time 1 s Rest phase 10 s Rise time 2 s Contraction phase 10 s Fall time 2 s Rest phase 50 s Rise time 2 s Contraction phase 10 s Fall time 2 s Rest phase 40 s...
Rise time 2 s Contraction phase 10 s Fall time 2 s Rest phase 20 s Rise time 2 s Contraction phase 10 s Fall time 2 s Rest phase 10 s Etius – User Manual Page 36 / 60...
8.Definition and parameters Symbol Description Available parameters Rise time 3 s Contraction phase 15 s Fall time 3 s Rest phase 50 s Rise time 3 s Contraction phase 15 s Fall time 3 s Rest phase 30 s Amplifier operation CC - stabilized output current mode CV - stabilized output voltage...
Page 38
0,5 mA in the range of 0-10 mA, Maximum 1 mA in the range of 10-140 mA. • amplitude 0 – 140 V in CC mode, max. 140 mA Regulation: • 0.5 V in the range of 0-140 V. Etius – User Manual Page 38 / 60...
8.Definition and parameters Frequency modulation program FM determines in what time relations the base frequency modulation in the set up range occurs. The FM program describes the duration of lowest base frequency, time of increase to its highest value and time of falling to its lowest value. If SPECTRUM value is set up at 0 Hz, it is not possible to use frequency modulation program.
As opposed to the classic technique of generating interferential currents within patient’s tissue (frequency interference), with the Etius unit the internal modulation process has been transferred to the inside of the unit. This technology causes generation of interference field occupying much space (the therapeutic current passes larger tissue space than with the classic method), interference occurs even in case of not very precise electrode application, which simplifies the treatment method.
Page 41
8.Definition and parameters Parameters description: Symbol Description Available parameters Treatment time 1 - 60 minutes Diadynamic interferential current Shape of the Isoplanar interferential current current One-channel AMF current Carrier frequency Possible settings 2000 Hz, 4000 Hz, 6000 Hz, 8000 Hz, 10000 Hz Basic frequency Regulation in the range of 1 Hz - 100 Hz Frequency...
Warning means that such a setting is the reverse of commonly accepted way of marking polarity. Automatic polarization switch in the half performed treatment. Etius – User Manual Page 42 / 60...
8.Definition and parameters Symbol Description Available parameters Amplifier operation CC - stabilized output current mode CV - stabilized output voltage 0 – 100 mA in CC mode Regulation: • 0,5 mA in the range of 0-20 mA, Maximum 1 mA in the range of 20-100 mA, •...
Parameters description: Symbol Description Available parameters Treatment time 1 - 60 minutes positive Shape of the negative current alternating Pulse duration Regulation in the range of 1 - 500 ms Etius – User Manual Page 44 / 60...
Page 45
8.Definition and parameters Symbol Description Available parameters Basic frequency Regulation in the range of 0.3Hz - 500 Hz Frequency Regulation in the range of 5 Hz - 500 Hz spectrum Frequency modulation program is switched off Frequency rise time 3 s Frequency fall time 3 s Frequency rise time 1 s Hold time of maximum frequency 2 s...
200÷300 µs, f 5÷50 Hz – stimulation of neuromotoric units • • BURST – strong analgesic effect • SP-TENS – for spastic paralysis of the nervous muscle system Etius – User Manual Page 46 / 60...
Viscose electrode covers can be also washed in boiling water by the time of 1 minute, after that it is recommended to soak them in saline solution to improve conducting properties. Before immersion in the boiling water, it is recommended to soak viscose pads in the cool water. Etius – User Manual Page 50 / 60...
10. Maintenance, cleaning, disinfection If the viscose covers have a material tear or damage to the seams, replace them with new ones. NOTE: Used electrodes and viscose pads should be disposed of with hospital waste. 10.3 Checking the condition of interconnecting cables and electrodes 10.3.1 Checking of cable condition To check the electrode condition you may use the function available in setup mode as "Electrodes test”.
Switch the unit off and on once again. if the problem persists or Unit Error. frequently occurs, note down the error number and contact your service. Etius – User Manual Page 52 / 60...
10. Maintenance, cleaning, disinfection 10.5 Fuse replacement NOTE: Before proceeding to the further described operations isolate the unit from the mains supply! In case of burnt fuses, they must be replaced. Their parameters are given in chapter „Specification and accessories” and on the name plate. To replace fuses: Step Description...
RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the Etius unit is used exceeds the applicable RF compliance level above, the unit should be observed to verify normal operation.
Page 56
RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the Etius unit is used exceeds the applicable RF compliance level above, the unit should be observed to verify normal operation.
11. Specification and accessories 11.3 Standard accessories Name Quantity Etius - the controller A-UE-AST-EEWH Mains cable a) A-AE-AST-KPET2M_A or Patient’s cable: A-AE-AST-KPWPR2M_A a) Channel A b) A-AE-AST-KPET2M_B or b) Channel B A-AE-AST-KPWPR2M_B Electrodes 6x6 cm A-AE-AST-EL6060R or A-AE-AST-EL6060RV2 Electrodes 7,5x9 cm...
Catalogue number, symbol ISO 7000-2493 Operator’s manual, symbol ISO7000-1641 Disposal of used devices together with other waste is prohibited, complied with the requirements of WEEE Follow operating instructions, symbol ISO 7010-M002 Background color: blue Etius – User Manual Page 58 / 60...
Page 59
12. Appendix A Symbol Explanation No sitting, symbol ISO 7010-P018 Background color: white Circular band and slash: red Symbol or text: black No stepping on surface, symbol ISO 7010-P019 Background color: white Circular band and slash: red Symbol or text: black No pushing, symbol ISO 7010-P017 Background color: white Circular band and slash: red...
Page 60
12. Appendix A Etius – User Manual Page 60 / 60...