Manual Cleaning And Disinfection; Manual Cleaning - Hitachi UST-52124 Instruction Manual

Electronic sector probe
Table of Contents

Advertisement

MN1-5516 Rev. 10

5-5. Manual cleaning and disinfection

Prepare following items before manual cleaning and disinfection.
Probe
1) Detergent: ENZOL
material compatibility for this medical device.
2) Disinfectant: Cidex
compatibility for this medical device.
3) 2 tanks, 1 for cleaning and 1 for disinfection - optional: 1 additional tank for rinsing with deionized/
tap water. (sufficient size for immersion of the submergible part of the probe at full length)
4) Soft, fluff free cloth or single use towel
5) Personal protective equipment (gloves, water repellent protective skirt, face protection mask or
protective glasses see also instructions of the manufacturer for the detergent and the disinfectant)

5-5-1. Manual cleaning

Probe
1) The temperature of the detergent solution should be between 15-30 °C [59-86 °F], concentration is 1.6%.
Please note the minimum contact time of the detergent in the manufacturer's instruction.
If a differing detergent is used, please also consider the approved material compatibility for this
probe.
2) Immerge the submergible part of the probe (see figure) without connector into the detergent.
3) Wipe the submergible part of the probe under the surface of the detergent solution with a single-use,
fluff free soft cloth to remove all visible soil. Be sure that all grooves of the probe are implemented
during the cleaning process. If necessary use an appropriate cleaning brush for this purpose.
4) Wipe the non-submergible parts of the probe with a soft cloth dipped with a detergent.
5) Rinse the submergible part of the probe with running tap water for 1 minute.
6) Alternatively to step 5 suspend the submergible part of the probe in a tray filled with deionized
water/tap water for 5 min.
7) Visually check the outer surface of the probe for cleanness. If necessary, use magnifying glass for
visually check. If there is still soil visible, repeat all above steps.
/Cidezyme
(Johnson & Johnson, #2258) or another cleaning agent with approved
®
®
OPA (Johnson & Johnson, # 20391) or another disinfectant with approved material
®
-24-