Preparations For Use; Start Up Check; Visual Check; Verification Of Cleaning, Disinfection And Sterilization - Hitachi UST-52105 Instruction Manual

Electronic sector probe
Table of Contents

Advertisement

3. Preparations for Use

3-1. Start up check

3-1-1. Visual check

Visually check the probe tip, ultrasonic irradiation area, cable and connector.
If any holes, indentations, abrasion, cracks, deformation, looseness, discoloration, or other abnormalities
are found, do not use the equipment.

3-1-2. Verification of cleaning, disinfection and sterilization

Verify that cleaning, disinfection and sterilization are conducted according to the intended use.

3-1-3. Verification of operation

Connect to the ultrasound diagnostic instrument by following the instructions in section 4-2, "Connecting
to the ultrasound diagnostic instrument" and check that the selected probe match the sector display and
the displayed frequency and check the image for errors.
Remarks
For details on the displayed screens, see the documentation supplied with the ultrasound diagnostic
instrument.
If the probe is operated in still air, brightness on the top of the image may be non uniform, but this does
not affect the performance of the probe.
Warning
Be sure to preparations for use.
Using the probe without noticing an abnormal condition can result in injury to the operator or patient.
If an inspection finds an abnormal condition in the probe, immediately stop use and contact one of
our offices and/or distributor's offices listed on the back cover.
Caution
Do not use the probe if the selected probe and image do not match the frequency.
An incorrect acoustic output can result in burns or other injuries to the patient. Contact one of our
offices and/or distributor's offices listed on the back cover.
-13-
MN1-5064 Rev.16